If a calibration point had already been entered at 25mm (or approx. this value) a warning message
will be displayed. Press the key under OVERWRITE POINT to accept this new calibration point.
Press the ENTER key to return to the "Probe Calibration" screen.
Reset the calibration micrometer to 0.000mm.
Check that the measured probe is 0mm +/- 0.01mm. If the measured probe is not 0mm +/- 0.01mm,
enter a new calibration point, overwriting the previous point.
Reset the calibration micrometer to 25.000mm.
Check that the measured probe is 25mm +/- 0.01mm. If the measured probe is not 25mm +/- 0.01mm,
enter a new calibration point, overwriting the previous point.
Reset the calibration micrometer to 0.000mm.
Check that the measured probe is 0mm +/- 0.01mm. If the measured probe is not 0mm +/- 0.01mm,
enter a new calibration point, overwriting the previous point.
Set the calibration micrometer to 2.500mm.
Check that the measured probe is 2.5mm +/- 0.01mm. If the measured probe is not 2.5mm +/- 0.01mm,
enter a calibration point, overwriting the previous point if requested.
Perform the complete calibration by entering points at 2.5mm intervals, i.e. at 5mm, 7.5mm, 10mm,
12.5mm, 15mm, 17.5mm, 20mm, 22.5mm and a new calibration pointy at 25mm.
88 PET Plus User Manual