Ametek Davenport PET Plus User Manual page 87

Viscometer including polymer dryer and polymer grinder
Table of Contents

Advertisement

The true probe should automatically display 0.000mm.
If it does not show 0.000mm, press the key under CHANGE TRUE then enter 0mm.
The display will now show 0.000mm for the true probe.
Press the key under ACCEPT to correct the zero probe reading.
If a calibration point had already been entered at 0mm (or approx. this value) a warning message will
be displayed. Press the key under OVERWRITE POINT to accept this new calibration point.
Press the ENTER key to return to the "Probe Calibration" screen.
Set the calibration micrometer to 25.000mm.
Check that the measured probe is 25mm +/- 0.01mm.
If the measured probe is not 25mm +/- 0.01mm, press the key under CAL POINT to display the "cal
point" screen.
The true probe should automatically display 25.000mm.
If it does not show 25.000mm, press the key under CHANGE TRUE then enter 25mm.
The display will now show 25.000mm for the true probe.
Press the key under ACCEPT to correct the zero probe reading.
PET Plus User Manual
87