Technical Specifications - Ametek Davenport PET Plus User Manual

Viscometer including polymer dryer and polymer grinder
Table of Contents

Advertisement

17.0 TECHNICAL SPECIFICATIONS

IV Measurement
Range
Barrel
Dies
Heaters
Temperature Range
Temperature Control
Over temperature Device
Nitrogen Gas
Pressure Input
Working Pressure Range
Pressure Measurement
Mounting
Optional Computer
Probe Unit
Ancillary Equipment (optional)
Electrical Supply
Power Consumption
PET Plus
Polymer Dryer
Fuses
PET Plus
Polymer Dryer
Operating Temperature
Storage Temperature
Relative Humidity
Net Instrument Weights
PET Plus
Polymer Dryer
Polymer Grinder
Dimensions (HxWxD)
PET Plus
Polymer Dryer
Polymer Grinder
102 PET Plus User Manual
Minimum 0.45 using 0.015" Bore Die
Maximum 1.40 using 0.040" Bore Die
Insulated stainless steel barrel with replaceable stainless steel liner
One die is supplied with the equipment (size specified at time or
order) but additional dies may be purchased as required.
Dies are available with bores of 0.015", 0.020", 0.030" and 0.04"
3 x 250 W cartridge type symmetrically positioned for optimum
temperature control
50°C - 400°C
Solid state, electronic 3 term controller with set point adjustment
Secondary overheating safety limit, factory set to 400ºC
Oxygen free and less than 5ppm of water specification (White
Spot). Recommended 200 Bar capacity bottle
30 Bar (440 psi)
Maximum 35 Bar (510 psi)
7 - 30 Bar
Electronic pressure transducer
Bench mounting. Enough bench space will also be needed for the
optional computer, key board and possibly printer
PC compatible.
Optional software supplied on CD
Linear Voltage Displacement Transducer encased in non-magnetic
canister
Polymer Drier, Vacuum Pump, Polymer Grinder, Polymer Chip
Transfer Vessel, Industrial Grinder and Melt Pot
230V AC +10% 50/60Hz or
115V AC +10% 50/60Hz
Not adjustable.
Mains Power Must be specified at time of order.
800 VA (max)
1000 VA (max)
Mains In 230V AC, 2 x T5AH250V
Mains In 115V AC, 2 x T10AH250V
Mains In 230V AC, 2 x T10AH250V
Mains In 115V AC, 2 x T10AH250V
- 5°C to 35°C
-20°C to 55°C (must be substantially dry)
0 to 80% (non-condensing)
34kg
15kg
41kg
64cm high (94cm with probe fitted) 47cm wide, 42cm deep
45cm high (with tubes fitted), 36cm wide, 45cm deep
40cm high (52cm high when open), 40cm wide, 54cm deep

Advertisement

Table of Contents
loading