The test is performed in exactly the same way
as for a stand-alone system, i.e. the sample is
inserted into the barrel then the GO button is
pressed.
Note that the computer now requires no user
intervention until the end of the test.
NEXYGEN Plus PET will initially plot the barrel
temperature and nitrogen pressure and will then
plot the probe position, probe speed and instan-
taneous IV at the relevant times during the test.
At the end of the test, the "Line of Best Fit" will
be drawn through the IV values and the IV at
Zero will be calculated. The selected test results
will also be calculated and stored in the data
table.
To perform the next test, click on the NEW button at the left hand side of the graph.
The button will change to a green START TEST button and clicking on this button will perform the
next test.
When 2 or more tests have been performed in a batch, the graphs can be quickly displayed
sequentially by clicking on the LEFT or RIGHT buttons at the left hand side of the graph.
To view the data table, click on the RESULTS button at the left hand side of the graph or click on the
RESULTS tab at the bottom of the screen to data table will display the results as shown.
98 PET Plus User Manual