Setting The Timer - Ametek Davenport PET Plus User Manual

Viscometer including polymer dryer and polymer grinder
Table of Contents

Advertisement

10.6 SETTING THE TIMER

The timer can be set to a wide range of time
periods but its main use is provide a 50 minutes
or 55 minutes timing for drying polymer powder.
Check that the 3 controls are set as below:
n Mode = Top Right Screw = A
n Full Scale = Bottom Left Screw = 1
n Time Scales = Bottom Right Screw = h
A small terminal screwdriver is required to
adjust.
Rotate the relevant screw to set the unit as
shown below:
Set the dial to 0.82 to select 50 minutes.
Using a suitable container and laboratory balance, weigh out 3.6g of polymer powder. Pour the
sample into a sample drying tube (glass phial) then screw the cap onto the top of the tube. Tighten
the cap sufficiently to give a good seal between the silicone rubber gland and the glass tubing but do
not over tighten.
Turn the vacuum valve in the plastic vacuum tube to closed (lever is at 90 degrees to the vacuum
pipe).
A
1
h
Set the dial to 0.92 to select 55 minutes.
PET Plus User Manual
63

Advertisement

Table of Contents
loading

Table of Contents