9.3.2
Move displacer
The move displacer operation is optional and can be used to change the current position of
the displacer in order to perform the calibration steps more easily.
1. Navigate to: Setup → Calibration → Move displacer → Move distance
2. Input the relative moving distance for the Move distance parameter.
3. Select the Move down option or the Move up option
4. Select the Yes.
This completes move displacer commands procedure.