Element Position 0 To - Endress+Hauser Proservo NMS83 Operating Instructions Manual

Tank gauging
Hide thumbs Also See for Proservo NMS83:
Table of Contents

Advertisement

Proservo NMS83
Endress+Hauser
Lower diameter of stilling well
The dimension of D
has to be the larger value of the dimensions D
2
See the table below.
Concentric pipe
D
Dimension
D
Dimension
2
2x
(Example)
Example
>100.9 mm
68.1 mm
(3.97 in)
(2.68 in)
100.9 mm
(3.97 in)
Example: L
= 20 000 mm, d = 50 mm, s = 5.0, 28 m drum
2
Asymmetric pipe
D
Dimension
D
Dimension
2
2x
(Example)
Example
>84.5 mm
68.1 mm
(3.33 in)
(2.68 in)
84.5 mm
(3.33 in)
Example: L
= 20 000 mm, d = 50 mm, s = 5.0, 28 m drum
2
Recommendations for NMS8x mounting with a stilling well
Follow the recommendations for mounting NMS8x with a stilling well.
• Keep the pipe connection welds smooth.
• When drilling holes into the pipe, keep the interior surface of the holes clear of
metal chips and burrs.
• Keep the pipe as vertical as possible. Check using a plumb bob.
• Install the asymmetric pipe under the valve and fit the centers of the NMS8x and
the valve.
• Set the center of the lower part of the asymmetric pipe in the direction of the lateral
motion.
• Observe the recommendations as per API MPMS chapter 3.1B.
• Confirm grounding between NMS8x and the tank nozzle.
Description
Parameter
D
Calculated D
value
1
1
D
D
dimension when the displacer is
2b
2
in L
length
2
Description
Parameter
D
Calculated D
value
1
1
D
D
dimension that the displacer can
2b
2
pass through (nth groove)
Installation
and D
.
1
2b
Formula
= 2 x (|p (L
)| + d/2 + s)
2
Formula
= |p (L
)| + d/2 + s + D
/2
2
1
23