Proservo NMS83
Additional information
Calibration status
Navigation
Description
Additional information
Make low table
Navigation
Description
Selection
Factory setting
Additional information
Set low weight
Navigation
Description
User entry
Factory setting
Endress+Hauser
Read access
Write access
Setup → Calibration → Drum cal. → Status
Gives feedback on the latest status of the calibration process.
Read access
Write access
Setup → Calibration → Drum cal. → Make low table
For additional accuracy it is possible to perform a second drum calibration with low weight.
Choose ' Y es' or ' N o' to start/stop calibration.
• No
• Yes
No
Read access
Write access
Meaning of the options
• No: A second drum calibration is not performed.
• Yes: A second drum calibration with low weight will be performed.
Setup → Calibration → Drum cal. → Set low weight
Set weight for additional drum calibration sequence.
10 to 999.9 g
Dependent on the device version
Operator
-
Operator
-
Operator
Maintenance
Operating menu