About this document CYR52 About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious injury. Causes (/consequences) If necessary, Consequences of non- compliance (if applicable) ‣...
Intended use The CYR52 ultrasonic cleaning system is used to clean turbidity sensors in liquids that are installed in pipes. Use of the device for any purpose other than that described, poses a threat to the safety of people and of the entire measuring system and is therefore not permitted.
Basic safety instructions CYR52 Product safety 2.5.1 State-of-the-art technology The product is designed to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. The relevant regulations and international standards have been observed.
Incoming acceptance and product identification CYR52 Incoming acceptance and product identification Incoming acceptance Verify that the packaging is undamaged. Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved. Verify that the contents are undamaged.
Click the product overview. A new window opens. Here you fill information pertaining to your device, including the product documentation. 4.2.3 Manufacturer address Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 D-70839 Gerlingen Scope of delivery The scope of delivery comprises: •...
CYR52 Mounting Mounting the ultrasonic transducer CAUTION Substances in the sealant Irritation to eyes and skin! ‣ Do not inhale aerosols if used. ‣ Avoid contact with the eyes and skin. ‣ Wear protective gloves, protective goggles and protective clothing when using the sealant.
Mounting CYR52 Mounting the ultrasonic generator 5.3.1 Post mounting You require the post mounting kit (optional) to mount the unit on a pipe, post or railing (round or square, clamping range 20 to 60 mm (0.79 to 2.36 in)). A0033044 ...
Page 19
CYR52 Mounting A0033045 A0038625 13 Post mounting 14 Attach the device and click it into place Place the device on the mounting plate. Slide the device downwards in the guide on the mounting rail until it clicks into place.
Page 20
Mounting CYR52 A0025886 A0038628 16 Rail mounting 17 Attach the device and click it into place Place the device on the mounting plate. Slide the device downwards in the guide on the mounting rail until it clicks into place.
CYR52 Mounting A0027799 A0038631 20 Wall mounting 21 Attach the device and click it into place Place the device on the mounting plate. Slide the device downwards in the guide on the mounting rail until it clicks into place.
Electrical connection CYR52 Electrical connection WARNING Device is live! Incorrect connection may result in injury or death! ‣ The electrical connection may be performed only by an electrical technician. ‣ The electrical technician must have read and understood these Operating Instructions and must follow the instructions contained therein.
CYR52 Electrical connection Connect the ultrasonic cleaning system in the following order: Connect the control cable (not included in the scope of delivery, H03VV-F 2x0.75 is recommended) to terminals 1 and 2 at the ultrasonic generator. Connect the control cable to a free relay at the transmitter at terminals 41 and 42.
Commissioning CYR52 Commissioning Function check WARNING Incorrect connection, incorrect supply voltage Safety risks for staff and device malfunctions! ‣ Check that all connections have been established correctly in accordance with the wiring diagram. ‣ Ensure that the supply voltage matches the voltage indicated on the nameplate.
CYR52 Diagnostics and troubleshooting Diagnostics and troubleshooting General troubleshooting Error Possible cause Test Cleaning effect diminishes Ultrasonic cleaning system is Acoustic check (if the ultrasonic transducer hums defective during the cleaning process, the cleaning is working) The position of the ultrasonic...
Maintenance CYR52 Maintenance WARNING Process pressure and temperature, contamination, electrical voltage Risk of serious or fatal injury ‣ If the sensor has to be removed during maintenance work, avoid hazards posed by pressure, temperature and contamination. ‣ De-energize the device before opening it.
The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium.
Accessories CYR52 Accessories The following are the most important accessories available at the time this documentation was issued. ‣ For accessories not listed here, please contact your Service or Sales Center. CYY101 • Weather protection cover for field devices • Absolutely essential for field installation •...
CYR52 Technical data Technical data 12.1 Output 12.1.1 Operating frequency Operating frequency of the ultrasonic cleaning system: approx. 40 kHz 12.2 Power supply 12.2.1 Supply voltage Depends on the order version: • 230 VAC ± 10 %, 50/60 Hz • 115 VAC ± 10 %, 50/60 Hz 12.2.2...
Page 30
Technical data CYR52 12.4.2 Weight Ultrasonic transducer: 0.72 kg (1.59 lb), with 3 m (9.8 ft) cable Ultrasonic generator: 2.2 kg (4.85 lb) 12.4.3 Materials Ultrasonic transducer Cover: Floor: Aluminum Cable: TPE-U mix; 3xLi9Y 0.75; sheath: Ø 6.6 mm (0.26 in) Minimum bending radius: 66 mm (2.6 in)66 mm (2.6 inch) when cable can...