Linear approximation of the HyTD correction
The real amount of deformation varies non-linearly with the level due to the construction
of the tank. However, as the correction values are typically small compared to the
measured level, a simple straight line method can be used with good results.
H
71
Calculation of the HyTD correction
1
Linear correction according to "Deformation factor (→ 253)"
2
Real correction
3
Starting level (→ 252)
L
Measured level
H
HyTD correction value (→ 252)
Calculation of the HyTD correction
L
Measured level
L
Starting level
0
c
HyTD correction value
HyTD
D
Deformation factor
3
L
0
L < L
=
C
= 0
0
HyTD
L > L
=
C
= - (L - L ) x D
0
HyTD