Proservo NMS83
Set level
Navigation
Description
User entry
Factory setting
Additional information
Level source
Navigation
Description
Selection
Factory setting
Additional information
High stop level
Navigation
Description
*
Visibility depends on order options or device settings
Endress+Hauser
Setup → Set level
If the level measured by the device does not match the actual level obtained by a manual
dip, enter the correct level into this parameter.
0 to 100 000 mm
0 mm
Read access
Write access
The device adjusts the Empty parameter (→ 170) according to the entered value, such
that the measured level will match the actual level.
Setup → Level source
Defines the source of the level value.
• No input value
• HART device 1 ... 15 level
*
• Level SR
*
• Level
*
• Displacer position
• AIO B1-3 value
• AIO C1-3 value
• AIP B4-8 value
• AIP C4-8 value
Dependent on the device version
Read access
Write access
Setup → High stop level
Position of the displacer high stop as measured from defined zero position (tank bottom or
datum plate).
Operator
Maintenance
Operator
Maintenance
Operating menu
171
Need help?
Do you have a question about the Proservo NMS83 and is the answer not in the manual?