Proservo NMS83

Set level

Navigation
Description
User entry
Factory setting
Additional information
Level source
Navigation
Description
Selection
Factory setting
Additional information
High stop level
Navigation
Description
*
Visibility depends on order options or device settings
Endress+Hauser
Setup → Set level
If the level measured by the device does not match the actual level obtained by a manual
dip, enter the correct level into this parameter.
0 to 100 000 mm
0 mm
Read access
Write access
The device adjusts the Empty parameter (→  170) according to the entered value, such
that the measured level will match the actual level.
 Setup → Level source
Defines the source of the level value.
• No input value
• HART device 1 ... 15 level
*
• Level SR
*
• Level
*
• Displacer position
• AIO B1-3 value
• AIO C1-3 value
• AIP B4-8 value
• AIP C4-8 value
Dependent on the device version
Read access
Write access
 Setup → High stop level
Position of the displacer high stop as measured from defined zero position (tank bottom or
datum plate).
Operator
Maintenance
Operator
Maintenance
Operating menu
171