7.8.9 Entering the offset of the external temperature probe

If a deviation is found during the check using a calibrated reference thermometer probe, then the offset (the additive
part of the characteristic) of the external measurement chain can be adjusted with the following function. The refer-
ence thermometer must be dipped nearly by the external temperature probe into the consumer bath according to the
details on the calibration certificate.
Master
3x
to
MNEnu
2x
to
CAL..
2x
Command
intern Pt100
extern Pt100
Pump
Menu
56 / 123
End
T
T
set
fix
CAL E
 Call the offset adjustment for the external temperatures probe
CAL E.
 Continue as described in ( 7.8.7) for the internal temperature probe.
Calibration
Default
End
T
T
set
Proline heating thermostat
 Confirm the control dialog on the right with
or cancel with
 Return to the measurement window with
End or
Calibration
 Open the device parameter menu via the soft key
Menu .
 With the cursor keys continue to:
Settings
extern Pt100
 The adjacent window appears.
 Confirm selection with
 Continue as described in ( 7.8.7) for the inter-
nal temperature probe.
fix
.
.
Calibration
Calibration
.
.
02/2022