Lauda RP 845 Operating Instructions Manual

Lauda RP 845 Operating Instructions Manual

Low-temperature thermostats with smartcool system
Table of Contents

Advertisement

Operating instructions

RP 845, RP 855, RP 870, RP 890, RP 1290
Valid from series: 07-0001 (see item 9.4)
YACE0072 / 08/07
replaced 04-0001
from softwareversion of Control system (Master) 1.46
from softwareversion of Protection system (Master) 1.33
from softwareversion of Operating system (Command) 1.69
from softwareversion of Cooling system 1.44
from softwareversion of Analogue interface 1.33
from softwareversion of RS232/485-Module 1.38
from softwareversion of contact I/0 module 1.38
from softwareversion of solenoid valve 1.37
PROLINE
Low-temperature Thermostats
with SmartCool System
RP 1840, RP 1845, RP 3530
LAUDA DR. R. WOBSER
GMBH & CO. KG
P.O.Box 1251
97912 Lauda-Koenigshofen
Germany
Phone: (+49) 9343/ 503-0
Fax: (+49) 9343/ 503-222
E-Mail info@lauda.de
Internet http://www.lauda.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RP 845 and is the answer not in the manual?

Questions and answers

Summary of Contents for Lauda RP 845

  • Page 1: Operating Instructions

    Operating instructions PROLINE Low-temperature Thermostats with SmartCool System RP 845, RP 855, RP 870, RP 890, RP 1290 RP 1840, RP 1845, RP 3530 Valid from series: 07-0001 (see item 9.4) YACE0072 / 08/07 LAUDA DR. R. WOBSER GMBH & CO. KG replaced 04-0001 P.O.Box 1251...
  • Page 3: Safety Notes

    Proline Low-temperature Thermostats Safety notes Before operating the equipment, please read carefully all the instructions and safety notes in Section 1. If you have any questions, please phone us! Follow the instructions on setting up, operation etc. This is the only way to avoid incorrect operation of the equipment and to ensure full warranty protection.
  • Page 4: Table Of Contents

    Proline Low-temperature Thermostats Contents SAFETY INFORMATION............................8 ..........................8 ENERAL SAFETY INFORMATION ........................... 8 THER SAFETY INFORMATION BRIEF OPERATING INSTRUCTIONS ........................9 ........................... 10 ENU STRUCTURE ASTER ..........................11 ENU STRUCTURE OMMAND CONTROLS AND FUNCTIONAL ELEMENTS ..................... 12 UNIT DESCRIPTION ..............................15 ..........................
  • Page 5 Proline Low-temperature Thermostats LMnet B/<P- RDp-- 7.6.12 Submenu (Master): Displaying refrigerating system settings......51 ..............................52 MPORTANT SETTINGS 7.7.1 Temperature setpoint setting...........................52 7.7.2 Displaying the actual external temperature....................54 7.7.3 Setting pump power or standby........................55 7.7.4 Activating external control ..........................56 7.7.5 Current consumption from the mains......................57 7.7.6 Setting the date and time (Command)......................59 7.7.7...
  • Page 6 Proline Low-temperature Thermostats 8.4.1 Menu structure Analogue module (Master)....................110 ..............................111 ONTACT MODULE 8.5.1 Contact module LRZ 915 with three inputs and three outputs ..............111 8.5.2 Namur-Contact module LRZ 914 with only one input and one output ............112 8.5.3 Menu structure contact module (Master) .....................
  • Page 7 Proline Low-temperature Thermostats Explanation of signs: Danger: This sign is used where there may be injury to personnel if a recommendation is not followed accurately or is disregarded. Note: Here special attention is drawn to some aspect. May include reference to danger. Reference Refers to other information in different sections.
  • Page 8: Safety Information

    Proline Low-temperature Thermostats Safety information General safety information A laboratory thermostat heats and circulates liquids according to specified parameters. This involves hazards due to high temperatures, fire and general hazards due to the application of electrical energy. The user is largely protected by the application of relevant standards. Further hazard sources may arise due to the type of tempering medium, e.g.
  • Page 9: Brief Operating Instructions

    Proline Low-temperature Thermostats Brief operating instructions These brief instructions shall give you the possibility to operate the unit quickly. For safe operation of the unit, it is necessary to read carefully all the instructions and safety notes! 1. Assemble unit and add items as appropriate ( 6.1).
  • Page 10: Menu Structure: Master

    Proline Low-temperature Thermostats Menu structure: Master Display flashes. Set setpoint Setpoint Bath temperature with . Display °C °C 1%21 is accepted afer 4 sec. or proceed immediately with the key: ⇒ Section 7.7.1 Pump power.1-8/Standby.0 External temperature °C °C 1%/1 Display flashes.
  • Page 11: Menu Structure: Command

    Proline Low-temperature Thermostats Menu structure: Command Pump Level intern Pt100 Calibration Menu extern Pt100 Default Calibration All Modules all default Works settings Master Resolution All default Command Device status Only control par. int. Cool Display data Only control par. ext. Other connected Basic settings Only miscellaneous...
  • Page 12: Controls And Functional Elements

    Proline Low-temperature Thermostats Controls and functional elements Mobile Command Console (see page 14) Master control panel (see page 14) Mains switch Pump connection at side: Suction nozzle (return to bath) Bypass valve (see illustration on next page) Side pump connection: Pump outflow, pressure output Bath cover Recessed grip...
  • Page 13 Covers for the two module slots Connection socket 10S for the external Pt100 temperature probe Connection socket 70S (LAUDA internal bus (LiBus)) for bus suitable for unit and to which the refrigerating lower section, the integrated electrically heated cover plate and Command...
  • Page 14 Proline Low-temperature Thermostats Error signal (red LED blinking) Control element: Master Bath controlled by external temperature source (geen LED lits) The temperature of an external source is displayed (EXT is lit green) Cooler active (blue LED is lit) Heater active (yellow LED is lit) Select and Enter keys Overtemperature set point to check or set T...
  • Page 15: Unit Description

    50 mm²/s should not be exceeded. The temperature control is the best with 30 mm²/s and lower viscosity. With small bath coolers (e.g. RP 845) power level 1 to 3 is practicable. Unit description YACE0072 / 21.08.07...
  • Page 16: Materials

    Proline Low-temperature Thermostats For operation as a circulating thermostat with an external load, a higher power level is practicable to maintain the temperature difference low, among other things also with higher temperatures in conjunction with oils as heat carrier liquid. The pump connections on the unit are fitted with M16 x 1 thread.
  • Page 17: Programmer And Ramp Function

    • For the connection of an external Pt100 temperature sensor (10S). • Two sockets (70S) for the connection of components via the LAUDA equipment bus (cooling section, Command Console, external solenoid valve, etc.). Command Version: The Command unit is equipped as standard with the following sockets: •...
  • Page 18: Cooling Unit

    LRZ 915, but only one output and one input on each of two DIN sockets. Coupling socket 3-pole, LAUDA Order No. EQD 047 and coupling plug 3-pole, LAUDA Order No. EQS 048. Further details can be found in section 8.5.
  • Page 19: Unpacking

    After unpacking, firstly check the device and accessories for any damage in transit. If, contrary to expectations, there is visible damage to the unit, the shippers or the postal service must be immediately informed, so that an investigation can be made. Please also inform the LAUDA Service Centre (Contact 9.4.).
  • Page 20: Preparation

    − Only RP 890 and RP 1290: In case of disturbing dewing, connect the heating of the bath bridge. Plug the supplied T-piece adaptor cable for the LAUDA device bus into the 70S socket and secure it. Connect it with the connection cable of the bath bridge heating.
  • Page 21: Filling And Draining

    Proline Low-temperature Thermostats Filling and draining Filling − Close the drain cock. − Carefully remove all residues of the previous heat carrier liquid (blow dry and remove screw plugs!). − Maximum filling level is up to 10 mm below the top edge of the bath. Overfilling leads to VW@pl0/2 the display of the warning Section 7.14.4).
  • Page 22: Heat Carrier Liquids And Hose Connections

    (flash point 128 °C). Check the mixture ratio with a hydrometer. − At devices with nickel-plated evaporator (RP 845, RP 855, RP 890, RP 1290 and RP 1845) you must not use acid, aqueous heat carrier liquids resp. detergents (ph-value < 7).
  • Page 23 Proline Low-temperature Thermostats Hose connections a) Elastomer hoses Hose type Internal width Temperature Field of Order number Ø mm range °C application For all heat carrier EPDM hose liquids except 10...120 RKJ 111 uninsulated Ultra 350 and mineral oils For all heat carrier EPDM hose liquids except 10...120...
  • Page 24: Connecting External Loads

    Proline Low-temperature Thermostats Connecting external loads Operation as circulating thermostat − When used as circulation thermostat, care for shortest hose connections with largest inner diameter as possible. This gives the best flow. − Push hose with 11-12 mm internal width onto hose olive or connect metal hoses ( 6.3) to pump connectors.
  • Page 25: Starting Up

    Proline Low-temperature Thermostats Starting up Mains connection Compare the rating on the nameplate (back of control head and behind the front panel) with the mains voltage. The device is according to EMC standard DIN EN 61326-1 Class B. − Connect unit only to sockets with a protective earth conductor (PE). −...
  • Page 26 Proline Low-temperature Thermostats KDTDK − Display for (low level) appears when the bath has too Level alarm little liquid. KDTDK °C − Red LED above the fault triangle flashes. − Find cause of fault and, where necessary, top up missing heat carrier liquid ( Section 6.3).
  • Page 27: Switching Off / Standby

    Proline Low-temperature Thermostats Switching off / standby Switching off: Set mains switch to position 0. With operation at temperatures below 0° Celsius only switch off the device via standby, because otherwise severe condensation and dew formation can occur. Standby operation: Use the key on the command console or by selecting the pump level Zero on the master display.
  • Page 28 Proline Low-temperature Thermostats − The following always applies: After termination of the relevant settings, they are accepted automatically after approx. 4 s − the setting is accepted immediately with the Enter key. − Fault signal: Flashing red Alarm LED and acoustic signal. −...
  • Page 29 Proline Low-temperature Thermostats Brightness Contrast Display Brightness The brightness and contrast can be set on the Sounds Master Contrast Command Console: Sounds Command − The works setting can be changed via Language Settings Basic settings Master Mode Display Brightness Contrast Autostart Current Consumpt.
  • Page 30 Proline Low-temperature Thermostats 3. Super window with seven items of °C °C 55,3 information: Y(%) − T , current bath temperature. − T , setpoint. °C − T , current temperature on external probe (if connected). − Overtemperature cut-off point T Pump °C −...
  • Page 31: Changing Window Information (Command Console)

    Proline Low-temperature Thermostats 7.4.2 Changing window information (Command Console) Command Display data You can adapt the information displayed by °C Level your Command Console to your requirements. For example, if you have not connected any temperature probe, you can exchange it in the standard setting of the °C normal window for the maximum temperature (safety cut-off).
  • Page 32: Locking The Keyboard

    Proline Low-temperature Thermostats 7.4.3 Locking the keyboard The keyboards of the Master and the Command Console can be locked independently of one another. This is especially advantageous when the thermostat is positioned in another room and the Command Console is used as a remote control device. Then the Master keyboard can be locked to prevent unintentional adjustment.
  • Page 33 Proline Low-temperature Thermostats Command Locking: − Press and then and hold Locking keyboard pressed simultaneously for three seconds. − The locking window appears. − Hold both keys pressed until the progress bar is completely filled. − Then the display skips back to the previously set Screen mode.
  • Page 34: Master" Menu Structure

    Proline Low-temperature Thermostats "Master" menu structure Bsp Max. current consumpt. TDp Software version Section 7.7.5 Section 9.2.2 R Setpoint resolution Rlp^G Serial number Hi Word. Section 7.8.1 Section 9.2.3 1%21 Rr@pr Opern. w. mains fail. Rlp^K Serial number Lo Word. Section 7.8.2 Actual bath temperature Section 9.2.3...
  • Page 35: Command" Menu Structure

    Proline Low-temperature Thermostats "Command" menu structure Pump Level intern Pt100 Calibration Menu extern Pt100 Default Calibration All Modules all default Works settings Master Resolution All default Command Device status Only control par. int. Cool Display data Only control par. ext. Other connected Basic settings Only miscellaneous...
  • Page 36: Basic Settings And Branching To Submenus (Master)

    Proline Low-temperature Thermostats 7.6.1 Basic settings and branching to submenus (Master) Master; Main level Display flashes. Set setpoint Setpoint Bath temperature with . Display °C °C 1%21 is accepted after 4 sec. or proceed immediately with the key: ⇒ Section 7.7.1 Pump power 1-8/Standby.0 External temperature °C...
  • Page 37: Submenu Lmdkt (Master): Branching To Further Submenus

    Proline Low-temperature Thermostats LMDkt 7.6.2 Submenu (Master): Branching to further submenus Reduce current consumpt., Further settings Device parameters resolution, start after mains °C LMDkt °C O@pA- fail, works settings, sounds, device type. ⇒ Section 7.6.4 Parameters for internal Control parameters controller (bath controller) &...
  • Page 38: Submenu Lmnet (Master): Configuration Of Modules

    Proline Low-temperature Thermostats LMnet 7.6.3 Submenu (Master): Configuration of modules Module/component-specific possible settings are only shown, when the hardware is connected. Protection system version, Configure modules Protection system characteristics and settings °C of protection system. LMnet °C RDC-- ⇒ Section 7.6.9 Indicates, that a Command Activate Command Console is connected.
  • Page 39: Submenu Lmdkt O@Pa- (Master): Configuration Of Device Parameters

    Proline Low-temperature Thermostats LMDkt O@pA 7.6.4 Submenu - (Master): Configuration of device parameters Display flashes. Set Device parameters with between Max. current consumpt. 10.0A ... 16.0A. °C O@pA- °C ⇒ Section 7.7.5 Display flashes. Set Resolution with 0.1=0.1°C or 0.01=0.01°C. °C R </0 ⇒...
  • Page 40: Submenu Lmdkt O0C-- (Master): Setting Internal Control Parameters

    Proline Low-temperature Thermostats LMDkt O0c-- 7.6.5 Submenu (Master): Setting internal control parameters You only obtain the following possible settings when the control variable is set to "Internal" (⇒ Section 7.7.4. See also 7.6.6.) Display flashes. Set Set control parameters with Proportional range [K] °C O0c--...
  • Page 41: Submenu Lmdkt O0C

    Proline Low-temperature Thermostats LMDkt O0c-- 7.6.6 Submenu (Master): Setting external control parameters You only obtain the following possible settings when the control variable is set to "External" (⇒ Section 7.7.4). See also 7.6.5. Display flashes. Set Set control parameters with Prop.rng.master contr.[ ] °C O0c--...
  • Page 42 Proline Low-temperature Thermostats LMDkt O0c-- (Master): Setting external control Continuation of submenu parameters Return to the previous page * Important: The windows for setting the Display flashes. Changeover additional parameters for the longer tne, tve and tde from external control for the Master than =Automatic to =Manual...
  • Page 43: Submenu Lmdkt B@K

    Proline Low-temperature Thermostats LMDkt B@K-- 7.6.7 Submenu (Master): Calibrating internal and external Pt100 temperature probe measuring chains (adjustment) longer than Display flashes. Use Calibrate temp. probe Offset internal probe to enter value °C B@K-- °C B@K 0 of reference thermometer. 3 sec.
  • Page 44: Submenu Lmdkt Rgmy \ (Master): Displaying Internal Values Of The Thermostat

    Proline Low-temperature Thermostats LMDkt RgmY[ 7.6.8 Submenu (Master): Displaying internal values of the thermostat Flashing display: Displays Software version Software version of control °C RgmY[ °C system. Flashing display: Serial no. Hi Serie no. of thermostat (Hi °C Rlp^G word). Flashing display: Serial no.
  • Page 45 Proline Low-temperature Thermostats RgmY[ Continuation of submenu : Displaying internal values of the thermostat Return to previous page Flashing display: Ext. temp. analog interf. Ext.temperature on analogue input [°C]. °C rD@l@ Flashing display: Ext. temp. serial interf. Ext. temperature on °C rDRDp RS232/ 485 input [°C].
  • Page 46 Proline Low-temperature Thermostats RgmY[ Continuation of submenu : Displaying internal values of thermostat Flashing display: 24V supply voltage Actual value of 24V supply Return to previous voltage [V]. °C page Flashing display: Pump power Momentary pump power °C OT O consumption [W].
  • Page 47: Submenu Lmnet Rdc-- (Master): Safety System Settings

    Proline Low-temperature Thermostats LMnet RDC-- 7.6.9 Submenu (Master): Safety system settings The safety modul takes care of all safets related tasks. It is integrated in the Master and it cannot be plugged in and out as other modules can. Some settings as adjusting the acoustic alarm level of RG/VW messages or to view parameters ( ) can be accessed over the Master settings as well.
  • Page 48: Submenu

    Proline Low-temperature Thermostats LMnet B/<P- 7.6.10 Submenu (Master): Refrigerating system settings Display flashes. Set with . The compressor may be switched from Refrigerating system Configure refrig. system automatic Bm @ to manuel B/<P- °C / / Bm 0 mode. °C Show various device values: Show values T01, T02, PT3, SM pos, ect.
  • Page 49: Submenu Lmnet B/

    Proline Low-temperature Thermostats LMnet B/<P- RgmY[ 7.6.11 Submenu (Master): Displaying refrigerating system settings Display flashes: Display Software version Software version of the RgmY[ °C °C refrigerater system. Display flashes: Serial no. Hi Serial no. of the thermostat °C Rlp^G (Hi-Word). Display flashes: Serial no.
  • Page 50 Proline Low-temperature Thermostats LMnet B/<P- RgmY[ Continuation of submenu (Master): Displaying refrigerating system settings Return to the previous page Display flashes: Display temperature temperature tc [°C]. °C r rb Display flashes: Display temperature temperature LP [°C]. °C r KO Display rotations of the fan Fan rotation in [min °C...
  • Page 51: B/

    Proline Low-temperature Thermostats LMnet B/<P- RDp-- 7.6.12 Submenu (Master): Displaying refrigerating system settings Display flashes. Set Servicemenüs* with Low-level warning °C RDp-- °C Range: 1 … 8. Flushing Start flushing. °C E /EE * Changes by Acclimation running Start acclimation running. authorised °C @c@Or...
  • Page 52: Important Settings

    Proline Low-temperature Thermostats Important settings 7.7.1 Temperature setpoint setting The setpoint is the temperature, which the thermostat should reach and maintain constant. Master (main level) − Press key until (Setpoint) appears. − Press, display flashes. − Enter the setpoint with the two keys ( Section 7.4.1 General key functions and pilot lamps).
  • Page 53 Proline Low-temperature Thermostats Command or T − or the soft key opens the setpoint window. − 123.45 is the setpoint, which is still active. The upper and lower limit temperatures are displayed (device-specific values). Enter new setpoint: There are three different possible entry methods: 123, 1.
  • Page 54: Displaying The Actual External Temperature

    Proline Low-temperature Thermostats − Select desired position with the cursor keys (black background). Enter new setpoint: − With the soft key Edit open the 123. window shown on the left. − Enter fixed temperature setpoint as described above and accept into the list Min: -40.00°C Max:202.00°C with or cancel with...
  • Page 55: Setting Pump Power Or Standby

    Proline Low-temperature Thermostats Master − Switches to the actual-value display of the external temperature probe (or to the actual value received from an interface module − EXT is lit in green next to the row of figures. − If no external Pt100 probe is connected, ,,,,, /1%/1 °C...
  • Page 56: Activating External Control

    Proline Low-temperature Thermostats Command Pump Level − Open the device parameter menu via the Pump Level Level 8 soft key Menu . Level 7 − Change from Pump Pump Level Level 6 Level 5 using Level 4 − With you enter the illustrated Level 3 window.
  • Page 57: Current Consumption From The Mains

    Proline Low-temperature Thermostats Change temperature probe source: − for the internal probe, − only when an external probe is connected, − only when an analog module is connected and configured, − only when a serial module is connected and is continuously receiving actual values from a PC.
  • Page 58 Proline Low-temperature Thermostats Master − Call current consumption LMDls O@p@ − The present setting is displayed. °C 0&/ − The current consumption is shown flashing: e.g. − Set the required maximum current consumption (in A). − Display flashes 4 s Wait 4 seconds or new value is automatically accepted, or −...
  • Page 59: Setting The Date And Time (Command)

    Proline Low-temperature Thermostats 7.7.6 Setting the date and time (Command) Command Clock Time Date − Open the device parameter menu via the Pump Set time Settings Set date soft key Menu . Graph Timer 1 − With the cursor keys continue to: Clock Timer 2 Clock...
  • Page 60: Special Settings

    Proline Low-temperature Thermostats − Select the desired resolution with Resolution 0,01 − Accept selection with 0,001 or End , or quit the window with without making changes. Help Menu Special settings 7.8.1 Setpoint resolution This function enables the resolution of the setpoint T to be increased from the standard value of 0.1 °C to 0.01 °C.
  • Page 61 Proline Low-temperature Thermostats Rr@pr Master Rr@pr − Call the start option LMDls O@p@ − The start mode can be changed here. °C Rr@pr @srm LM@l − The display flashes. @srm − , when operation is be restored automatically again after an interruption.
  • Page 62: Defining Temperature Limits

    Proline Low-temperature Thermostats 7.8.3 Defining temperature limits With this function, it is possible to define a minimum and a maximum temperature in which the thermostat controls. By reaching the temperature limits, a warning appears. In this way setpoint input can be prevented which may damage the heat carrier medium or the apparatus. For example, if water is used as the heat carrier liquid, +95°C would be practicable as the maximum temperature and +5°C as the minimum temperature.
  • Page 63: Setpoint Offset Operating Mode

    Proline Low-temperature Thermostats − Enter the desired limit temperature. − Accept the change with Lower limit (T il) -50. − or quit the window with without making changes. Min: -50.0 °C Max: 301.0 °C Help Menu Screen 7.8.4 Setpoint offset operating mode With this function it is possible to apply an offset value to the temperature provided by the external temperature probe or a module and then to use it as the setpoint.
  • Page 64 Proline Low-temperature Thermostats Command Offset source Setpoint offset − Open the device parameter menu via the Offset source extern Pt100 Setpoint offset soft key Menu . RS232 − With the cursor keys continue to Control Setpoint offset Offset source − indicates that the setpoint offset is currently deactivated.
  • Page 65: Restoring Works Settings

    Proline Low-temperature Thermostats 7.8.5 Restoring works settings Master If you would like to restore all the works settings except the control parameters and the probe calibrations − call the works settings LMDls O@p@ − is displayed. °C − Press longer than 3 sec. cmlD −...
  • Page 66 Proline Low-temperature Thermostats − Under Master you have the choice between: − all default , then all Master settings are reset, − only control para. int. for the internal control parameters, − only control para. ext. similar for external, − only miscellaneous which resets setpoint, pump level, maximum...
  • Page 67: Setting The Volume Of The Acoustic Signals

    Proline Low-temperature Thermostats 7.8.6 Setting the volume of the acoustic signals The LAUDA Proline Thermostats signal alarms as a dual-tone acoustic signal and warnings as a continuous tone. @schm Master @schm − Call the volume setting 7.6.4). LMDls O@p@ @schm −...
  • Page 68: Entering The Offset Of The Internal Temperature Probe

    7.8.7 Entering the offset of the internal temperature probe If, during checking with a calibrated reference thermometer (e.g. from the LAUDA DigiCal Series) a deviation is found, then the offset (i.e. the additive part of the characteristic) of the internal measuring chain can be adjusted with the following function.
  • Page 69: Restoring The Works Setting Of The Internal Temperature-Probe Offset

    Proline Low-temperature Thermostats − The temperature measurement device shows the true temperature value (with Temperature value of the ref. glass thermometers consider the temp. measurement device: correction where applicable!). − Change the display in the adjacent window to the true value with cursor or soft keys and accept with or End , Min: -50.0°C...
  • Page 70: Entering The Offset Of The External Temperature Probe

    7.8.9 Entering the offset of the external temperature probe If a deviation is found during the check using a calibrated reference thermometer, e.g. from the LAUDA DigiCal Series, then the offset (the additive part of the characteristic) of the external measurement chain can be adjusted with the following function.
  • Page 71: Restoring The Works Setting Of The External Temperature-Probe Offset

    Proline Low-temperature Thermostats 7.8.10 Restoring the works setting of the external temperature-probe offset If the offset has been misadjusted unintentionally, the works setting can be restored with this function. cDE D Master − Call the offset works setting for the external temperature probe cDE D LMDls −...
  • Page 72: Graphical Display Of Temperature Measurements (Command)

    Proline Low-temperature Thermostats Graphical display of temperature measurements (Command) Command Screen and Graph 25.00 25.01 25.02 − Press the soft key Screen a number of T °C 27.00 times as required until the graph recorder window appears. 26.00 − With the soft key Graph you enter the 25.00 menu for the configuration of the graph...
  • Page 73 Proline Low-temperature Thermostats Temp. scale defines how the scaling is to be Mode Temp. min 22.00 carried out: Displayed value Temp. max 27.00 − automatic , by the program, or Legend − manual in that you yourself define the Sample Time limits with the next menu point.
  • Page 74: Programmer (Pgm Only Command)

    Proline Low-temperature Thermostats 7.10 Programmer (PGM only Command) Almost any temperature/time profile can be created with the programmer. A desired bath temperature can be approached as quickly as possible or via a defined ramp. Furthermore, the pump level and the behavior of the switching outputs can be defined.
  • Page 75 Proline Low-temperature Thermostats Each program begins with the segment "Start". It defines at which temperature Segment 1 is to continue the program. It is not possible to specify a time for the Start segment. For thermostats without cooling ability, the start temperature must be selected higher than the bath temperature, which prevails before the program start.
  • Page 76: Selecting And Starting The Program (Start, Hold, Stop)

    Proline Low-temperature Thermostats 7.10.2 Selecting and starting the program (Start, Hold, Stop) Here you will learn how to select and start a program that has already been created. If no program has been created 7.10.4 Creating or modifying a program (Edit). Command Programmer Program 1...
  • Page 77: Interrupting, Continuing Or Terminating The Program (Hold, Continue, Stop)

    Proline Low-temperature Thermostats 7.10.3 Interrupting, continuing or terminating the program (Hold, Continue, Stop) Command Programmer Program 1 Status − After a program has been started by Hold Status pressing the key, the command Stop Edit options Hold Stop are shown. Loops Graph −...
  • Page 78: Creating Or Modifying A Program (Edit)

    Proline Low-temperature Thermostats 7.10.4 Creating or modifying a program (Edit) Here, there are the following functions: • Entry of a program. • Display of the program data of a saved program and modification of the segment data. • Insertion or appending of a new segment. •...
  • Page 79 Proline Low-temperature Thermostats − In the "Start" line enter in the field "T end T end °C Time [h:m] Tolerance °C" the temperature at which the sequence is to start (default value is Start 30.00°C ------- 3.00°C 30°C). A time entry is not possible in the 30.00°C 00:30 3.00°C...
  • Page 80 Proline Low-temperature Thermostats − A new segment is produced by moving the cell with the black background to a blank line with the cursor keys and then End of segment temperature: pressing the soft key Insert . The values of the cell located above it are automatically copied.
  • Page 81 Proline Low-temperature Thermostats − If the field in the column "Pump" has a Level 8 Pump level black background, the entry mode for the Level 7 Pump level is obtained by pressing the Level 6 Level 5 − With Level 4 select Pump Level 1 –...
  • Page 82: Defining The Number Of Program Loops (Loops)

    Proline Low-temperature Thermostats 7.10.5 Defining the number of program loops (Loops) Command Programmer Program1 Loops − If required, programs can be looped Status many times. Edit − With Loops access the menu Loops Graph Info − Select the number of desired program loops.
  • Page 83: Obtaining Information On A Program (Info)

    Proline Low-temperature Thermostats − The display of the programmed T °C 27.00 temperature curve can be quit with End . 26.00 25.00 24.00 23.00 Help Menu 7.10.7 Obtaining information on a program (Info) Command Programmer Program1 Info − Continue with Info Segments Status...
  • Page 84: Ramp Function

    Proline Low-temperature Thermostats 7.11 Ramp function With the ramp function, temperature changes over any time period can be conveniently entered. This is especially advantageous with very low temperature changes (e.g. 0.1°C/ day). Example: From the current outflow temperature (e.g. 242.4 °C) 200°C of cooling is to occur over 5 days. Then the temperature change is entered as 200°C and the time as 5 days.
  • Page 85: Timer Function (Command)

    Proline Low-temperature Thermostats 7.12 Timer function (Command) Using the timer function, the thermostat can carry out an action at a certain time or after a certain waiting period. The actions are: switching on the thermostat, entering the standby mode or one of the 5 programs in the programmer.
  • Page 86: Control Parameters

    Proline Low-temperature Thermostats − Weekplan Arrange takes you to the Weekplan window shown on the left. Time Action Time Action − Using the cursor keys Monday 07:30 Start 17:00 ------- select the field, which is to be filled in. Tuesday 10:00 Prog.4 17:00 ------- −...
  • Page 87: Internal Control Variable (Integral Measurement Probe)

    Proline Low-temperature Thermostats 7.13.1 Internal control variable (integral measurement probe) Only read further here if you have no external temperature probe connected (and activated according to Section 7.7.4 as control variable). O0c-- Master − As shown in the menu structure ( 7.6.5 ), the parameters for the internal control variable can be set.
  • Page 88: External Control Variable (External Measurement Probe)

    Proline Low-temperature Thermostats 7.13.1.1 Proven settings for control parameters and pump (integral measurement probe) Instrument Heat carrier Liquid Pump Level Type RP 845 Water RP 845 Ethanol RP 845 Ethanol RP 855 Ethanol RP 855 Ethanol RP 855 Water RP 855...
  • Page 89 Proline Low-temperature Thermostats Parameter Slave controller (P-controller): − Sl proportional range: = Xpf in kelvin. − Tve, Tde, Prop_E changeover to Auto/ Man °C automatic or D LM = manual. Works setting is automatic. These three parameters should only be modified by experienced control technicians! Hold pressed for 3 seconds, then make changes.
  • Page 90 Description Volu- Hose- Prop_E ment carrier lenght Level Type liquid double wall glass vessel RP 845 Water double wall glass vessel RP 855 Water double wall stainless RP 855 Water steel vessel with water double wall glass vessel RP 855 Ethanol 15.0...
  • Page 91: Alarms, Warnings And Errors

    Minimum +40°C thermostatic liquid − Tools to watch the time behavior: Graph mode of the Command console, LAUDA Wintherm PC- program. 7.14 Alarms, Warnings and Errors The SelfCheck Assistant of your Proline Thermostat monitors more than 50 device parameters and triggers alarms, warnings or errors as appropriate.
  • Page 92 Proline Low-temperature Thermostats Setting the overtemperature cut-off: Recommended setting: 5°C above desired bath temperature. Caution!! The overheat switch-off point T is controlled by a system functioning independently of the bath control. Setting of the nominal and T temperature, however, can be limited via the functions T 7.8.3) independently of T Overtemp.
  • Page 93: Low-Level Alarm And Low-Level Checking

    − If irregularities arise during the checking of the safety devices, switch off the unit immediately and pull out the mains plug. − Have the equipment checked by LAUDA Service. Command Low-level alarm! −...
  • Page 94: High-Level Settings

    Proline Low-temperature Thermostats 7.14.3 High-level settings Different reactions can be chosen when the level sensor detects the height of the heat carrier liquid level. Depending on the setup, heat carrier liquid or operation conditions, one of the following settings may be suitable: Setting Master Command...
  • Page 95: High-Level Warning Or Alarm

    Proline Low-temperature Thermostats 7.14.4 High-level warning or alarm − Acoustic warning signal sounds for 3 seconds when the liquid level rises so far that the uppermost switching point of the level sensor 3 Sec. has been reached. Or in case the warning function as described in 7.14.3 was chosen: −...
  • Page 96: Pump-Motor Supervision: Overload Or Blockage

    Proline Low-temperature Thermostats 7.14.5 Pump-motor supervision: Overload or blockage The SelfCheck Assistant monitors the VarioFlex Pump: 1. Alarm sounds as dual-tone signal for pump-motor overload or blockage. aK/B 2. Display of signals blockage. Pump alarm °C aK/B 3. The red LED above the fault triangle flashes.
  • Page 97: Fault List „Alarms And Warnings

    Proline Low-temperature Thermostats 7.14.7 Fault list „Alarms and Warnings“ Alarms Message Meaning OsKDT Pump too fast (low level) KDTDK Low level alarm in the level sensor rDLMO Overtemperature (t > tmax) aK/B Pump blocked (no rotation) BE@0K Command connection interrupt Temperature signal of external Pt100 missing Temperature signal of analogue input missing Temperature signal of serial port missing...
  • Page 98 Proline Low-temperature Thermostats Warnings in the “Command-Display” Warnings from “Cooling system” Message Meaning Message Meaning VW1/0 Overflow of CAN receipt VW2/0 Overflow of CAN receipt VW1/1 Watchdog-Reset VW2/1 Watchdog-Reset RTC Voltage drop recognized: Battery failure sm.stell_min still not determined Adaption run VW1/2 VW2/2 necessary...
  • Page 99 Proline Low-temperature Thermostats Warnings from “Contact I/0-Module” Warnings from “Solenoid valve” Code 7, 8, 9XX) Message Meaning Message Meaning Overflow of CAN receipt VW5/0 VW6/0 Overflow of CAN receipt Watchdog-Reset VW5/1 VW6/1 Watchdog-Reset Software version of control system too old Software version of control system too old VW50/ VW60/...
  • Page 100: Rs 232 Interface

    Proline Low-temperature Thermostats 7.15 RS 232 Interface 7.15.1 Connecting cables and interface test RS 232 Computer Thermostat Signal 9-pin sub-D-socket 25-pin sub-D-socket 9-pin sub-D-socket Signal R x D T x D T x D R x D Signal Ground Signal Ground with hardware handshake: For connecting a thermostat to the PC use 1:1 cable and not a null- modem cable! without hardware handshake: the computer / PC must be set to the operating mode “without hard...
  • Page 101: Connecting Cable Rs 485

    Proline Low-temperature Thermostats Example: Transfer of setpoint 30,5°C to the thermostat Computer Thermostat “OUT_SP_00_30.5“CRLF “OK“CRLF 7.15.3 Connecting cable RS 485 Thermostat 9-pin sub-D-socket Data Data A (-) SG (Signal Ground) optional Data B (+) − Use screened connecting cables. − Connect screen to connector case. −...
  • Page 102: Write Commands (Data Commands To The Thermostat)

    Proline Low-temperature Thermostats Example: Transfer of setpoint 30.5°C to the thermostat with address 15. Computer Thermostat “A015_OUT_SP_00_30.5“CR “A015_OK“CR 7.15.5 Write commands (Data commands to the thermostat) Command Explanation OUT_PV_05_XXX.XX External temperature to be set through the interface. OUT_SP_00_XXX.XX Setpoint transfer with up to 3 places before the decimal point and up to 2 places behind.
  • Page 103: Read Commands (Data Requested From The Thermostat)

    Proline Low-temperature Thermostats Command Explanation segments receive this tolerance setting. − For ”_“ use also ” ” (blank character). − Response from thermostat ”OK“ or in case of error ” ERR_X“ (RS 485 interface e.g. “A015_OK” or in case of error ”A015_ERR_X”.). Permitted data formats: -XXX.XX -XXX.X...
  • Page 104 Proline Low-temperature Thermostats Command Explanation 0 = make-contact open/ 1 = make-contact closed. IN_DO_02 State of Contact output 2: 0 = make-contact open/ 1 = make-contact closed. IN_DO_03 State of Contact output 3: 0 = make-contact open/ 1 = make-contact closed. IN_MODE_00 Keys Master: 0 = free / 1 = inhibited.
  • Page 105: Error Messages

    ® programmed with the aid of the National Instruments program development tool LABVIEW (http://sine.ni.com/apps/we/nioc.vp?cid=1381&lang=US). In order to make program operation possible on the RS 232/ RS 485 interface, LAUDA provides drivers ® specially designed for LABVIEW which can be downloaded free of charge under www.lauda.de/spec-...
  • Page 106: Interfaces - Modules

    Proline Low-temperature Thermostats Interfaces - Modules Installing of modules The master and command can be supplemented with further interface modules, which are simply inserted at the back of the control head into two module slots. − Touch the earthed bath cover of the Proline thermostat to discharge any electrostatic charge.
  • Page 107: Menu Structure For All Modules (Only Command)

    Proline Low-temperature Thermostats Menu structure for all modules (only Command) All existing menu points are illustrated. However, the Command Console masks out menu points, which cannot be executed. Further information can be found in the following sections. RS232 Menu RS485 2400 Setpoint temperature 4800...
  • Page 108: Serial Interfaces Rs232 / 485

    RS232 / 485 Interface Module (order no. LRZ 913) with 9-pole SUB-D socket. Electrically isolated by optocoupler. With the LAUDA instruction set essentially compatible to the Ecoline and Integral Series. The RS232 interface can be connected directly to the PC with a 1:1 through-contact cable (order no.
  • Page 109: Analogue Module

    Proline Low-temperature Thermostats Analogue module The analogue module (order no. LRZ 912) has 2 inputs and 2 outputs, which are brought out on a 6- pole DIN socket to Namur Recommendation (NE28). The inputs and outputs can be set independently as 4…20 mA, o…20 mA or 0…10V interface.
  • Page 110: Menu Structure Analogue Module (Master)

    Proline Low-temperature Thermostats 8.4.1 Menu structure Analogue module (Master) All existing menu points are illustrated. However, the Master unit masks out menu points which cannot be executed. TRD / or0: Activate (0); LM rR : Take setpoint TDp : Display module software 1%21 deactivate (/) input /output temperature from input...
  • Page 111: Contact Module

    Proline Low-temperature Thermostats Contact module 8.5.1 Contact module LRZ 915 with three inputs and three outputs Contact module Cat. no. LRZ 915) on 15 pole SUB-D socket. With three relay contact outputs (changeover, max. 30V/ 0.2A) and three binary inputs for control via external voltage-free contacts. The following functions are made available by the inputs: Set fault with function: Master: E @K@ or Command: Fault...
  • Page 112: Namur-Contact Module Lrz 914 With Only One Input And One Output

    Proline Low-temperature Thermostats 8.5.2 Namur-Contact module LRZ 914 with only one input and one output Contact module (Cat. no. LRZ 914) with connector to NAMUR NE28. Functionality as LRZ 915, but only one output and one input on each of two DIN sockets. Contact module LRZ 914;...
  • Page 113: Menu Structure Contact Module (Master)

    Proline Low-temperature Thermostats 8.5.3 Menu structure contact module (Master) All existing dialogue boxes are illustrated. The Master unit however masks out commands, which cannot be executed. E XXX E /EE TDp : Display module software 1%21 Select input function Input switched off version Actual bath temperature E @K@ Alarm function on this...
  • Page 114: Maintenance

    Proline Low-temperature Thermostats Maintenance Cleaning Withdraw the equipment mains plug before cleaning. Cleaning can be carried out with water to which a few drops of surfactant (washing-up liquid) have been added and using a damp cloth. No water must enter the control section. Carry out appropriate decontamination if hazardous material is spilt on or in the equipment.
  • Page 115: Device Data

    Proline Low-temperature Thermostats 9.2.4 Device data Master LMDkt RgmVW Section 7.6.8 − Various device datas are displayed. Command Device data Settings Device status T ext Pt 25.70 Tint Device data Display T ext analog ---.-- Mains U(%) 100.74 T ext serial ---.-- Mains frequ.
  • Page 116: Servicing Repair And Disposal Informations

    − Withdraw the mains plug before all service and repair work. − Only specialists must carry out repairs in the control section. 9.3.1 Servicing LAUDA Thermostats largely require no service. If the heat carrier liquid becomes contaminated, it should be replaced Section 6.2.
  • Page 117: Repair Informations

    Proline Low-temperature Thermostats 9.3.2 Repair information If you need to send in a unit for repair, it is essential to first contact the LAUDA Service Centre 9.4. − When sending in the unit, ensure that it is carefully and properly packed. LAUDA cannot be held liable for any damage caused by improper packing.
  • Page 118: Help Desk And Ordering Replacement Parts

    Your contact for service and support: LAUDA Service Centre Telephone: +49 9343 / 503-236 (English and German) E-Mail service@lauda.de We are available any time for your queries, suggestions and criticism. LAUDA DR. R. WOBSER GMBH & CO. KG P.O. Box 1251 97912 Lauda-Koenigshofen Germany...
  • Page 119: Accessories

    8.5.1. Relays module with 1 input and 1 output NAMUR NE28 functionality 8.5.2. LRZ 914 channel. T-piece adapter cable for the LAUDA For the connection of further LiBus internal bus (LiBus) components (with heating thermostats two EKS 073 LiBus connections are not occupied and one with cooling thermostats).
  • Page 120 RP 1845; 4 frames fit in P 26. UG 073 ID . Application frame for 20 tubes, diam. 14- 1 frame fits in P 8, (P 12), RP 845, RP 855, UG 076 18 mm 80mm ID . RP 870, RP 890.
  • Page 121: Technical Data

    Proline Low-temperature Thermostats Technical data The figures have been determined according to DIN 12876. RP 845 RP 855 RP 870 RP 890 RP 1290 RP 1840 RP 1845 RP 3530 RP 845C RP 855C RP 870C RP 890C RP1290C RP1840C...
  • Page 122 Proline Low-temperature Thermostats RP 845 RP 855 RP 870 RP 890 RP 1290 RP 1840 RP 1845 RP 3530 RP 845C RP 855C RP 870C RP 890C RP1290C RP1840C RP1845C RP3530C Bath depth / usable depth 200 / 180 250 / 230 Height to top of bath Overall dims.
  • Page 123 Proline Low-temperature Thermostats Pump characteristics measured with water Technical data YACE0072 / 21.08.07...
  • Page 124 Proline Low-temperature Thermostats Cooling curves Cooling curves; Bath closed, heat carrier liquid ethanol, Pump Level 3; Time in minutes; Temperature in °C Technical data YACE0072 21.08.07...
  • Page 125: Index

    Proline Low-temperature Thermostats Index Acoustic signals......... 67 Decimal-point key ......28 Hazard sources........8 Activate standby ........ 55 cDE works settings......65 Hazards ..........8 cDE 0 Offset, works setting....69 Activating external control ....56 Heat carrier liquid cDE D Offset, works settg....71 Alarms ..........
  • Page 126 Proline Low-temperature Thermostats install ...........106 Switching off........27 Modules ..........17 R setpoint resolution......60 R@ED keyboard locking ......32 Safety functions .........91 T end °C ..........79 Nameplate ..........25 Safety information........8 Technical data ........121 Safety notes..........3 rDLMO Overtemperature alarm...92 Safety system ........16 Temperature ........59 Screen displays........29 display resolution......59 Offset adjustment .......68...
  • Page 127 / Personne responsable Hiermit bestätigen wir, daß nachfolgend aufgeführtes LAUDA-Gerät (Daten vom Typenschild): We herewith confirm that the following LAUDA-equipment (see label): Par la présente nous confirmons que l’appareil LAUDA (voir plaque signalétique): Serien-Nr. Type / Type Serial no. / No. de série:...