Activating External Control - Lauda PROLINE P 5 C Operating Instructions Manual

Hide thumbs Also See for PROLINE P 5 C:
Table of Contents

Advertisement

Command
Pump level
Pump
Menu

7.7.4 Activating external control

An external temperature probe can be connected to the Proline Thermostats. How this is done is explained in Section
7.7.2. If the set point temperature is to be controlled using this sensor instead of the internal sensor, the setting can be
made here.
Furthermore, control can also occur based on the signal from the analog or serial module ( 4.8).
Master
and 2 x
Con 1
42 / 123
Level 8
Level 7
Level 6
Level 5
Level 4
Level 3
Level 2
Level 1
End
T
+
Please exercise caution when thermostat is in standby mode.
The following settings/ actions may start the thermostat unintentionally from the standby mode:
A previously activated timer mode ( 7.12), because a started time continues to run.
 "Start" command via interfaces ( 8).
Con
Call the source selection for the control
 The momentary setting for the source is displayed,
 here
°C
from the internal temperature probe.
 The source display flashes.
Proline heating thermostat
Pump level
 Open the device parameter menu via the soft
 Change from
 With
 Select another pump level with
 or quit the window with
T
set
fix
Standby activation
 Standby activation
1
for internal, because control takes place using the temperature signal
key
Menu .
Pump
or
you enter the illustrated win-
dow.
Level 5
is active.
confirm with
or End ,
changes.
(Pump and heater are deactivated when the
green LED in the lower part of the key is lit).
Con
.
Pump level
using
.
or
and
without making any
02/2022