7.8.3 Defining temperature limits

With this function it is possible to define a minimum and a maximum temperature in which the thermostat controls. By
reaching the temperature limits, a warning appears. In this way setpoint input can be prevented which may damage the
heat transfer liquid or the apparatus. For example, if water is used as the heat transfer liquid, 95 °C would be practica-
ble as the maximum temperature and 5 °C as the minimum temperature.
Master
3x
to
MNEnu
to
3x
PArA
tih
tiL
or
Wait 4 seconds or
Command
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Pump
Menu
48 / 123
tih
 Call the start option
or with 4x
 The maximum temperature can be changed here.
°C
 The minimum temperature can be changed here.
°C
 The maximum temperature (minimum temperature) is displayed flashing.
 Set the required limit temperature.
 Display flashes 4 seconds  new value is automatically accepted, or
 value is immediately accepted with the Enter key.
T
il (min)
T
302,0 °C
ih (max)
End
T
set
Proline heating thermostat
tiL
and
.
tih
(maximum temperature),
tiL
(minimum temperature).
Limits
 Open the device parameter menu via the soft
-50,0 °C
key
 With the cursor keys continue to:
Limits
 The minimum and maximum temperatures are
displayed.
 Select the limit to be changed with
T
and confirm with
fix
Menu .
.
T il (min)
is currently active.
.
or
02/2022