Fault List "Alarms And Warnings - Lauda PROLINE P 5 C Operating Instructions Manual

Hide thumbs Also See for PROLINE P 5 C:
Table of Contents

Advertisement

7.14.7 Fault list "Alarms and Warnings"

Alarms
Message
Pump too fast (low level)
PuLEU
Low level alarm in the level sensor
LEUEL
Overtemperature (t
tEMNP
Pump blocked (no rotation)
bL0C
Command connection interrupt
CFA1L
Temperature signal of external Pt100 missing
AL
1
Temperature signal of analogue input missing
AL
2
Temperature signal of serial port missing
AL
3
Analogue module: Current input 1 interrupted
AL
4
Analogue module: Current input 2 interrupted
AL
5
Protection system: High bath level
AL
6
Error digital input (from V 1.30 on)
AL
7
Refill fail
AL
8
Warnings from "Master-Display"
Message
Overflow of CAN receipt
WX
1
Watchdog-Reset
WX
2
til-limitation active
WX
3
tih-limitation active
WX
4
Heatsink temperature
WX
5
Software version of protection system too old
WX 11
Software version of operating system too old
WX 12
Software version of analogue Interface too old
WX 14
Software version of RS 232 too old
WX 15
Software version of contact I/0 module too old
WX 16
Software version of Valve 0 too old
WX 17
Software version of Valve 1 too old
WX 18
Software version of Valve 2 too old
WX 19
Software version of Valve 3 too old
WX 20
Software version of Pump 0 too old
WX 21
Software version of Pump 1 too old
WX 22
Software version of Pump 2 too old
WX 23
Software version of Pump 3 too old
WX 24
86 / 123
Meaning
> tmax)
Meaning
Proline heating thermostat
Warnings from "Safety system"
Message
Overflow of CAN receipt
WX101
Watchdog-Reset
WX102
Close to bath overflow
WX103
Bath level is approaching switch off level or is out of
WX104
optional range
Heater 1 break
WX105
Heater 2 break
WX106
Heater 3 break
WX107
Software version of control system too old
WX110
Software version of operating system too old
WX112
Software version of analogue interface too old
WX114
Software version of RS 232 too old
WX115
Software version of contact I/0 module too old
WX116
Software version of Valve 0 too old
WX117
Software version of Valve 1 too old
WX118
Software version of Valve 2 too old
WX119
Software version of Valve 3 too old
WX120
Software version of Pump 0 too old
WX121
Software version of Pump 1 too old
WX122
Software version of Pump 2 too old
WX123
Software version of Pump 3 too old
WX124
Meaning
02/2022