7.8.8 Restoring the works setting of the internal temperature-probe offset

If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
3x
to
MNEnu
2x
to
CAL..
1x
dEF 1
3 seconds
Command
intern Pt100
extern Pt100
Pump
Menu
x
02/2022
20,
15
Max: 302,00
2
3
4
7
Screen
8
T
9
set
dEF 1
 Call offset works setting for the internal temperature probe
dEF 1
dEF 1
°C
 The works setting is accepted after pressing for 3 seconds and
displayed.
Calibration
Default
End
T
set
Proline heating thermostat
 The temperature measurement device shows
the true temperature value (with glass ther-
mometers consider the correction where appli-
cable!).
 Change the display in the adjacent window to
the true value with cursor or soft keys and ac-
cept with
 or quit the window with
5
changes.
T
0
fix
,
is displayed.
Default
 Open the device parameter menu via the soft
key
 With the cursor keys continue to:
Settings
intern Pt100
 The window shown adjacent appears.
 Confirm selection with
T
f
or End ,
without making
donE
Menu .
Calibration
Default
.
.
55 / 123
is

Advertisement

Table of Contents
loading

Table of Contents