Temperature value of the ref.
temp. measurement device:
Min: -50,0°C
1
Help
6
Menu
7.8.8 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
3x
to
MNEnu
2x
to
CAL..
1x
dEF 1
3 seconds
Command
intern Pt100
extern Pt100
Pump
Menu
x
02/2022
20,
15
Max: 302,00
2
3
4
7
Screen
8
T
9
set
dEF 1
Call offset works setting for the internal temperature probe
dEF 1
dEF 1
°C
The works setting is accepted after pressing for 3 seconds and
displayed.
Calibration
Default
End
T
set
Proline heating thermostat
The temperature measurement device shows
the true temperature value (with glass ther-
mometers consider the correction where appli-
cable!).
Change the display in the adjacent window to
the true value with cursor or soft keys and ac-
cept with
or quit the window with
5
changes.
T
0
fix
,
is displayed.
Default
Open the device parameter menu via the soft
key
With the cursor keys continue to:
Settings
intern Pt100
The window shown adjacent appears.
Confirm selection with
T
f
or End ,
without making
donE
Menu .
Calibration
Default
.
.
55 / 123
is