8.3.2 Protocol RS 232

CR = Carriage Return (Hex: 0D)
LF = Line Feed (Hex: 0A)
Example:

8.3.3 Connecting cable RS 485

Pin
1
5
6
92 / 123
 The interface operates with 1 stop bit, no parity bit and 8 data bits.
 Transfer rate either: 2400, 4800, 9600 (factory setting) or 19200 baud as selected.
 The RS 232 interface can be operated with or without hardware handshake, (RTS/CTS).
 The command from the computer must be terminated with CR, CRLF, or LFCR.
 The response of the thermostat is always terminated with CRLF.
 After each command sent to the thermostat, it is necessary to wait for the reply before
sending another command. This ensures that the sequencing of inquiries and answers is
clear.
Transfer of setpoint 30,5 °C to the thermostat
Computer
„OUT_SP_00_30.5"CRLF
Thermostat
9-pin sub-D-socket
Data
Data A (-)
SG (Signal Ground)
optional
Data B (+)
 Use screened connecting cables.
 Connect screen to connector case.
 The connections are galvanically isolated from the rest of the electronics.
 Any pins not in use must not be connected!!
Proline heating thermostat
Thermostat
„OK"CRLF