Display resolution
Pump
Menu
7.8

Special settings

7.8.1 Setpoint resolution

This function enables the resolution of the setpoint T
(only Master).
Master
3x
to
MNEnu
to
1x
PArA
S =1
or
Wait 4 seconds or

7.8.2 Defining the type of start mode

Usually it is desirable that the thermostat carries on operating again after an interruption in the voltage supply. Howev-
er, if for safety reasons you do not wish this, you can insert an intervening manual activation step.
46 / 123
0,1
0,01
0,001
End
T
set
S
 Call setpoint resolution
 The current setting is displayed (here
°C
 The resolution display
=1
=01
 Display flashes 4 s  new value is automatically accepted, or
 value is immediately accepted with the Enter key.
Proline heating thermostat
 Select the desired resolution with
Accept selection with
or quit the window with
changes.
T
fix
to be increased from the standard value of 0.1 °C to 0.01 °C
set
S
.
=1
flashes.
for 0.1 °C.
for 0.01 °C.
or End ,
without making
=1
for the works setting of 0.1 °C).