7.14.4 High-level warning or alarm

3 Sec.

Level warning
WX103
Level alarm
AL
6
Command
84 / 123
 Acoustic warning signal sounds for 3 seconds when the liquid level rises so far
that the uppermost switching point of the level sensor has been reached.
Or in case the warning function as described in ( 7.14.3) was chosen:
 The acoustic signal with dual-tone sounds.
 Warning
much liquid.
°C
 The
In case the alarm function as described in 7.14.3 was chosen:
 The acoustic signal with dual-tone sounds.
°C
 Red LED
 Heater switches off on both poles,
 Pump unit is switched off electronically.
 Find the cause of the fault. Possible causes may be:
1. Volume expansion on heating.
2. Feed to an external vessel may be interrupted so that only return suction
3. Heat transfer liquid taking up moisture.
 At Alarm: Press Enter key. Warnings disappear automatically when the cause
is gone.
 Also; press this key if the unit has been switched off in the fault state. Warn-
ings disappear automatically when the cause is gone.
High-level warning/ alarm
 The display shows
Warning. To release press Enter key
Security 3 Level too high
or
Alarm AL 6: Level too high
unlocking is only possible on the Master control panel.
Proline heating thermostat
WXArn103
WXArn
flashes by turns with the numeral.
above the fault triangle
is possible.
(high level) appears when the bath contains too
flashes.
is shown in the display and signifies that