7.8.7 Entering the offset of the internal temperature probe

If, during checking with a calibrated reference thermometer probe, a deviation is found, then the offset (i.e. the additive
part of the characteristic) of the internal measuring chain can be adjusted with the following function. The reference
thermometer must be dipped into the bath according to the details on the calibration certificate.
Master
3x
to
MNEnu
2x
to
CAL..
CAL 1
3 sec
or
Command
intern Pt100
extern Pt100
Pump
Menu
54 / 123
CAL 1
 Call the offset adjustment of the internal temperature probe
CAL..
CAL 1
°C
 The actual value of the bath temperature is displayed flashing.
 Set the value which you have read from the calibrated reference measure-
ment device (with glass thermometers consider the correction where applica-
ble!).
 The entered value is accepted and
Calibration
Default
End
T
set
Proline heating thermostat
.
is displayed.
donE
Calibration
 Open the device parameter menu via the soft
key
 With the cursor keys continue to:
Settings
intern Pt100
 The window shown on the left appears.
 Confirm selection with
T
fix
is displayed.
Menu .
Calibration
Calibration
.
.
02/2022