Command
T
°C
set
25,
T
°C
ext
25,
Pump
Menu
7.7.3 Setting pump power or standby
With the Proline Varioflex pump, 8 pump levels are available with which the bath circulation, flow rate and pressure, the
noise generated and the mechanical heat input can be optimized. This is particularly advantageous with coolers. With
smaller coolers (e.g. P8) without an external load, Power Level 3 to 4 is practicable and sufficient.
Master
and 1 x
Pu
5
or
Wait 4 seconds or
02/2022
Level
4
00
T
°C
int
25,
01
Pump
3
02
Screen
T
set
Pu
Call pump power levels display
The current pump level is displayed (here
°C
The pump levels display flashes.
Select pump level (pump speed = pump power):
1
0
Display flashes 4 s new value is automatically accepted, or
value is immediately accepted with Enter key.
Proline heating thermostat
T
ext
Provided an external temperature probe is con-
nected, its value is displayed in the lower left part
of the standard and super windows (applies to the
works setting for the window partitioning).
External actual temperatures can also be read in
via interface modules. ( 8).
T
fix
8
to
for pump operation. Pump responds immediately!
activates the standby function (pump and heater are deactivated).
Pu
.
5
).
41 / 123