6.3
Filling and draining
Filling
Close the drain cock!
Carefully remove all residues of the previous heat transfer liquid (blow dry and remove screw
Maximum filling level is up to 10 mm below the top edge of the bath. Overfilling leads to the dis-
Best operation is with a level 20 – 80 mm below the top edge of the bath.
Low-level cut-off occurs at about 95 mm (at P 12 C approx. 215 mm) below the top edge of
Draining
Completely drain the bath, external consumers, accessories and hose connections and flush or clean them (e.g. with
new heat transfer liquid).
02/2022
plugs!).
WX103
play of the warning
the bath.
The units are designed for use with non-flammable and flammable liquids to DIN EN 61010-2-
010. Flammable heat transfer liquids ( Section 6.4) may only be used below the flash point.
When using thermal transfer oils note that they expand on heating (approx. 10 % per 100 K).
With enclosed external loads, the overall expansion takes place in the bath.
Ensure that with the connection of an external load, the liquid level does not drop impermissibly
due to filling the load. Top up with heat transfer liquid if necessary.
Set the upper and lower temperature limits ( 7.8.3) in accordance with the limits of the heat
transfer liquid in use.
Follow the regulations for the disposal of used heat transfer liquid.
Do not drain heat transfer liquid when it is hot or at bath temperatures below 0 °C!
Proline heating thermostat
( Section 7.14.4).
Switch off the thermostat, withdraw the mains plug!
Let out the heat transfer liquid through the drain cock; fit a
hose when doing this.
The drain cock is placed on the back of the heating thermo-
stats.
1
Drain cock
2
Drain Nozzle
23 / 123