7.7.5 Current consumption from the mains

If your mains fuse is rated below 16 A, the current consumption can be reduced in steps from 16 A to 10 A using this
function. Of course the maximum heating power of 3.5 kW is then also reduced accordingly. Take into account wheth-
er other loads are still connected to the fused circuit or whether your Proline Thermostat is the only load.
02/2022
Change temperature probe source:
1
for the internal probe,
EP
EA
ES
tual values from a PC.
 Display flashes 4 s  new value is automatically accepted, or
 value is immediately accepted with Enter key.
EP EA
 If
then the green
°C
nal temperature signal.
Intern
Extern Pt100
Analog module
Digital module
End
T
set
Proline heating thermostat
only when an external probe is connected,
only when an analog module is connected and configured,
only when a serial module is connected and is continuously receiving ac-
ES
or
has been selected,
LED indicates that the control has regulated to the exter-
Control variable
 Open the device parameter menu with the soft
key
 With the cursor keys, change further to:
Intern
 Select other control variables (only displayed
when present) with
T
fix
 or quit the window with
changes.
Menu .
Control
Control variable
is currently active.
or
or End ,
without making any
.
and confirm with
43 / 123

Advertisement

Table of Contents
loading

Table of Contents