7.13.3 Internal and external control parameter sets

If a thermostat is used for a number of applications, which always leads to a change of the control parameters, these
control parameters (up to 9 sets) can be saved in the thermostat and activated again as required.
Also saving is useful for finding the best control parameters; in this way external management of the control parame-
ters can be avoided.
There are 9 sets (each for internal and external sets of control parameters) saved at the factory.
In this menu the control parameters cannot be edited, they are only displayed.
With
Activate
With
Upload actual
With
Default
set by the customer are lost).
Command
Control Parameters
Control para. sets
Tv manual/auto
Correction limitation
Pump
Menu
Status
intern
extern
Pump
Menu
02/2022
the currently valid control parameters are used.
the actual ones are read in and saved (for later reuse).
the set of control parameters saved at the works is loaded again (in this case the control parameters
Set 1
Set 2
Set 3
Set 4
Set 5
Set 6
Set 7
Set 8
Set 9
End
T
set
Activate
Upload actual
Default
End
T
set
Proline heating thermostat
Control parameter sets
Open the device parameter menu via the soft
key
With the cursor keys continue to:
Control
Control para. sets
The adjacent window appears.
Set 1 to Set 9.
Select the desired set with
with
T
fix
In the setting window (see left) the selected
set is listed under
extern
Under
is activated,
is read in and
the set, who was saved at the factory, is re-
stored.
T
fix
Menu
.
Control Parameters
.
and confirm
.
intern
or
in the display.
Status
the previously selected set:
79 / 123