7.8.4 Setpoint offset operating mode
With this function it is possible to apply an offset value to the temperature provided by the external temperature probe
or a module and then to use it as the setpoint. The bath temperature can, for example, be operated at -25 °C below
the temperature of a reactor, which is being measured by the external temperature probe.
Master
3x
to
MNEnu
to
5x
PArA
SEtq.
r 0FF
Wait 4 seconds or
r0FFS
Wait 4 seconds or
02/2022
-50,
0
Max: 301,0 °C
2
3
4
7
Screen
8
T
9
set
SEtq.
SEtq.
source of the actual value can be entered.
Continue with
°C
The offset operating mode is currently deactivated.
To activate it press
°C
(external via analog module) or
Display flashes 4 s new value is automatically accepted, or
value is immediately accepted with the Enter key.
Continue with
°C
To activate the relative offset input press
or
Display flashes 4 s new value is automatically accepted, or
value is immediately accepted with the Enter key.
Proline heating thermostat
Enter the desired limit temperature.
Accept the change with
or quit the window with
changes.
5
T
0
fix
in the following submenu the relative setpoint (Set) and the
.
and with
. The left-hand window is displayed.
.
,
without making
EP
change to
(external Pt100),
ES
(external via serial module).
and change the value with
EA