8.3.6 Read commands (Data requested from the thermostat)

Command
IN_PV_00
IN_PV_01
IN_PV_03
IN_PV_04
IN_PV_05
IN_PV_10
IN_PV_13
IN_SP_00
IN_SP_01
IN_SP_02
IN_SP_03
IN_SP_04
IN_SP_05
IN_PAR_00
IN_PAR_01
IN_PAR_02
IN_PAR_03
IN_PAR_04
IN_PAR_05
IN_PAR_06
IN_PAR_07
IN_PAR_09
IN_PAR_10
IN_PAR_14
IN_PAR_15
IN_DI_01
IN_DI_02
IN_DI_03
IN_DO_01
IN_DO_02
IN_DO_03
02/2022
Read bath temperature (outflow temperature)
Indication of the controlled temperature (internal/external Pt/external Ana-
logue/external Serial)
Read external temperature TE (Pt100)
Read external temperature TE (Analogue input)
Read current bath level
Read bath temperature (outflow temperature) in 0.001 °C
Read external temperature TE (Pt100) in 0.001 °C
Read temperature setpoint
Read current pump power level
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC)
Read current overtemperature switch-off point
Read current outflow temperature limit TiH
Read current outflow temperature limit TiL
Read control parameter Xp
Read control parameter Tn (181 = OFF)
Read control parameterTv
Read control parameter Td
Read control parameter KpE
Read control parameter TnE (980 = OFF)
Read control parameter TvE (0 = OFF)
Read control parameter TdE
Read value of correction limitation
Read control parameter XpF
Read setpoint offset
Read control parameter PropE
State of contact input 1: 0 = open/ 1 = closed
State of contact input 2: 0 = open/ 1 = closed
State of contact input 3: 0 = open/ 1 = closed
State of Contact output 1:
0 = make-contact open/ 1 = make-contact closed
State of Contact output 2:
0 = make-contact open/ 1 = make-contact closed
State of Contact output 3:
0 = make-contact open/ 1 = make-contact closed
Proline heating thermostat
Explanation
95 / 123