Command
Control Parameters
Control parameter sets
Tv manual/auto
Correction limitation
Pump
Menu
7.13.2.1
Proven settings for control parameters and pump (external control):
Heat
Device
trans-
Field of application
type
fer
liquid
P 8
Water
Double wall glass vessel
P 8
Water
Double wall glass vessel
Double wall stainless
P 8
Water
steel vessel with water
7.13.2.2
Steps for setting the control parameters for external control
1.
Activate external control ( 7.7.4).
2.
Setting the slave controller:
2.1. Set parameters to
Check for thermostat type and change when necessary (P....) ( 9.2.1).
Choose heat transfer liquid with low viscosity and high thermal capacity.
Ranking: Water, ethanol, water-glycol, oil, Fluorinert®,
02/2022
Kpe
0,50
Tne
100
Tve (auto) 83
Tde (auto)
Xpf
Prop_E (a) 30
End
T
set
External consumer
Hose
Volume
length
[L]
2.5
2.5
0.7
auto
; Xpf see table ( 7.13.2.1) (empirical value) in dependence upon:
Proline heating thermostat
Control Parameters
Open the device parameter menu via the soft
key
Menu .
With the cursor keys continue to:
8,3
Control
4,0
trol Parameters
The adjacent window appears.
Parameter endings:
e = Master controller, f = Slave controller.
Where applicable change parameters marked
T
fix
with (auto) to manual input with
al/auto
.
Select the parameters to be changed with
and confirm with
Then change the value in the following settings
window and confirm with
Correction limitation
7.13.2).
Master controller
(External controller)
Kpe
Tne
Tve
EP
En
Eu
[m]
2x1
2.0
80
60
2x1
2.0
150
130
2x1
0.5
70
50
Control Parameters
.
Tv manu-
.
.
see introduction (
Slave controller
(Internal control-
ler)
Tde
Prop_E
Xpf
Ed
Eb
iP
5.0
30
4.0
5.0
30
3.0
5.0
30
3.0
77 / 123
Con-
Pump
level
5
5
5