7.8.9 Entering the offset of the external temperature probe

If a deviation is found during the check using a calibrated reference thermometer, e.g. from the LAUDA
DigiCal Series, then the offset (the additive part of the characteristic) of the external measurement chain
can be adjusted with the following function. The reference thermometer must be dipped nearly by the
external temperature probe into the consumer bath according to the details on the calibration certificate.
Master
MNEnu
3x
to
CAL..
2x
to
2x
Command
intern Pt100
extern Pt100
Pump
Menu
56
End
T
T
set
CAL E
 Call the offset adjustment for the external temperatures probe
CAL E.
 Continue as described in ( 7.8.7) for the internal temperature
probe.
Calibration
Default
End
T
T
set
Starting up
Proline Edition X heating thermostats
 Confirm the control dialog on the right
with
 Return to the measurement window with
End or
fix
Calibration
 Open the device parameter menu via the
soft key
 With the cursor keys continue to:
Settings
extern Pt100
 The adjacent window appears.
 Confirm selection with
 Continue as described in ( 7.8.7) for the
internal temperature probe.
fix
or cancel with
.
.
Menu .
Calibration
Calibration
.
.
YACE0099 / 10/9/2018