Entering The Offset Of The External Temperature Probe; Restoring The Works Setting Of The External Temperature-Probe Offset - Lauda PROLINE Kryomats RP 4050 C Operation Manual

Hide thumbs Also See for PROLINE Kryomats RP 4050 C:
Table of Contents

Advertisement

7.6.8 Entering the offset of the external temperature probe

If a deviation is found during the check using a calibrated reference thermometer, then the offset (the additive part of
the characteristic) of the external measurement chain can be adjusted with the following function. The reference
thermometer must be dipped nearly by the external temperature probe into the consumer bath according to the details
on the calibration certificate.
Command
intern Pt100
extern Pt100
Pump
Menu

7.6.9 Restoring the works setting of the external temperature-probe offset

If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
intern Pt100
extern Pt100
Pump
Menu
64 / 137
Calibration
Default
End
T
T
set
Calibration
Default
End
T
T
set
Proline Kryomats
Calibration
 Open the device parameter menu via the soft key
Menu .
 With the cursor keys continue to
Settings
Calibration
 The adjacent window appears.
 Confirm selection with
 Continue as described in ( 7.6.6) for the
internal temperature probe.
fix
Default
 Open the device parameter menu via the soft
key
Menu .
 With the cursor keys continue to
Settings
extern Pt100
 The adjacent window appears.
 Confirm selection with
 Continue as described in ( 7.6.7) for the
internal temperature probe.
fix
Calibration
extern Pt100
.
.
Calibration
Default
.
.