9.2.4 Device data
Master
Command
T ext Pt
T ext analog
T ext serial
T cont. head 39.80 Level
T heatsink
Pump Pow.
Pump rpm
Pumpe Cur.
Pump
Menu
9.2.5 Fault memory
For the analysis and localization of faults the Command version includes a fault memory in which up to 45 fault and
alarm messages are saved.
Command
No Source Code Type
10 Safety 2
9 Safety 4
8 Contro. 32
7 Contro. 3
6 Contro. 9
5 Contro. 3
4 Contro. 4
3 Contro. 5
Low level
Pump
Menu
102 / 123
25.70 Tint
---.-- Mains U(%) 103.74
---.-- Mains Frequ.
51.68 Low voltage 27.90
44.90 5V Supply
5460 Fan Voltage
1.68 Cur. cons.
End
T
set
Date
Alarm ------------ -----------
Warn. 28.08.03 15:32:02
Error 17.07.03 10:52:02
Warn. 06.06.03 11:15:11
Alarm 05.06.03 08:45:01
Alarm 01.06.03 17:58:22
Warn. 28.05.03 20:01:22
Warn. 27.05.03 07:58:00
End
T
set
Proline heating thermostat
MNEmv ShoWX (
Various device datas are displayed.
Device data
25,58
data
50
Text shows various actual temperatures in °C
6
from ext. Pt100 and the modules.
T cont. head and T heatsink are temperatures of
5.00
electronics in the Master in °C.
7.0
Pump power in Watts, pump speed in rpm, pump
2,84
current in ampere.
T
indicates the current internal bath temperature
int
in °C.
T
fix
Mains voltage in % of nominal and mains frequen-
cy in hertz.
Level indicates the liquid level in the internal bath.
Low Voltage of power transformer, of the 5 V
supply and fan voltage in volt.
Cur. cons.: Mains current consumption in ampere.
Errorstore
Settings
Time
Display
The last message is at the top.
Each message line can be marked with the cur-
sor keys. The message appears in plain text in
the footer.
Under "Source", the CAN node is displayed
which signaled the fault.
Code is the number, which in the Master is
shown in the display until the cause has been
T
fix
rectified.
Type: Alarm, Warning or Fault (Error).
Section 7.5)
Settings
Device status
Display
.
Device status
.
Device
Errorstore
02/2022