High-Level Settings; Master - Lauda PROLINE P 5 C Operating Instructions Manual

Hide thumbs Also See for PROLINE P 5 C:
Table of Contents

Advertisement

7.14.3 High-level settings

Different reactions can be chosen when the level sensor detects the height of the heat transfer liquid level. Depending
on the setup, heat transfer liquid or operation conditions, one of the following settings may be suitable:
Setting
No warning
Warning
Warning and
heater off
Alarm

Master

4 x
to
to
SED..
nHWX
Command
Overlevel handling none
Pump
Menu
02/2022
Master set-
Command set-
tings
tings
none
nHnon
Warning
nHWX
Warning + heat-
nHWXH
er off
Alarm
nHALA
nHWX
 Choose the menu for the high level reaction.
MNpfv
 The factory setting is
2x
 Press
 Choose a setting from the above list.
°C
Warning
Warn.+ Heater off
Alarm
End
T
set
Proline heating thermostat
Reaction and application recommendation
Select only when no safety sensitive application. I.e. water as
heat transfer liquid.
Acoustic and optical warning as long as the level goes down. This
is the factory setting.
Warning
and additional
down. Recommended for flammable heat transfer liquids with
much higher flash point and temperatures above 100 °C.
Alarm
switches off the pump and the heater until the alarm is
removed by pressing
mended for external loads and flammable liquids.
. That is a warning.
nHWX
to enter the setup mode. The current setting flashes.
Overlevel handling
 Open the device parameter menu via the soft key
Menu .
 With the cursor keys continue to:
Settings
 The shown window appears.
 Select the preferred parameter with
firm with
See introduction for details.
T
fix
heater off
as long as the level goes
on the Master keyboard. Recom-
Overlevel handling
.
.
and con-
83 / 123

Advertisement

Table of Contents