Edit Basic Settings; Change The Calibration Of The Temperature Probe - Lauda LOOP 100 Operation Manual

Thermo-electric circulation thermostat
Table of Contents

Advertisement

5.6

Edit basic settings

5.7

Change the calibration of the temperature probe

V3
1.
Select the Menu  Setup  Limits  Lower limit (Til) menu item.
2.
Enter the temperature limit value.
3.
Press [OK] to confirm the new value.
This chapter explains the basic settings.
Set temperature unit: Degrees Celsius or degrees Fahrenheit.
Set the baud rate of the serial interface.
Set the sound volume of the signal tone; loud, medium, quiet, and off.
Reduce or increase the display brightness.
Autostart: Setting [off] or [on]
Sometimes it is desirable for the device to resume operation after a
power failure. However, you can introduce a manual activation step
for safety reasons.
The maximum rotation speed of the ventilator in the device is limited
between 70 and 140 %.
With the[Fan Limit] at 100 %, under normal conditions (ambient tem-
perature 20 °C) the specified cooling capacity is reached.
If a higher cooling capacity is required or if higher ambient tempera-
tures are present, the maximum ventilation fan rotation speed can
be increased. This does, however, increase the noise issued by the
device.
Reducing the maximum ventilation fan rotation speed can cut down
on the operating noise of the device. This does, however, reduce the
cooling capacity of the device.
Increase the maximum ventilator fan rotation speed when the
device is operating in a warmer environment.
If the device has reached the control range, the ventilator fan rota-
tion speed will automatically be reduced, regardless of the setting
[Fan Limit].
If the device heats up, the rotation speed of the ventilator fan will be
automatically reduced, regardless of the setting [Fan Limit].
The menu languages English, German, French, Spanish, and Italian are
available for the device.
1.
Select the Menu  Setup  Basic setup menu item.
2.
Select the menu item for the value that you want to change.
3.
Press [OK] to confirm.
4.
Change the new value with the arrow buttons.
5.
Press [OK] to confirm.
The setting is changed.
A calibrated reference thermometer with the desired level of accu-
racy is necessary. Otherwise you should not change the calibration.
LOOP
23 / 41