Device Data; Fault Memory (Command) - Lauda RP 845 Operating Instructions Manual

Low-temperature thermostats with smartcool system
Table of Contents

Advertisement

Proline Low-temperature Thermostats

9.2.4 Device data

Master
Command
T ext Pt
T ext analog
T ext serial
T cont. head 39.80 Level
T heatsink
Pump pow.
Pump rpm
Pumpe cur.
Help
Menu

9.2.5 Fault memory (Command)

For the analysis and localization of faults the Command version includes a fault memory in which up to
45 fault and alarm messages are saved.
Command
No. Source Code Type Date
10 Safety
9 Safety
8 Contro. 32 Error
7 Contro. 3 Warn. 06.06.03 11:15:11
6 Contro. 9 Alarm 05.06.03 08:45:01
5 Contro. 3 Alarm 01.06.03 17:58:22
4 Contro. 4 Warn. 28.05.03 20:01:22
3 Contro. 5 Warn. 27.05.03 07:58:00
Low level
Help
Menu
YACE0072 / 21.08.07
25.70 Tint
---.-- Mains U(%) 100.74
---.-- Mains frequ.
51.68 Low voltage 27.90
44.90 5V supply
5460 Fan voltage
1.68 Cur. cons.
End
T
set
2 Alarm ------------ -----------
4 Warn. 28.08.03 15:32:02
17.07.0310.:52:02
End
T
set
LMDkt
Various device datas are displayed.
Device data
-8
50
− T ext shows various actual temperatures
4
in °C from ext. Pt100 and the modules.
− T cont. head and T heatsink are
5.00
temperatures of electronics in the Master
7.0
in °C.
2.84
− Pump power in Watts, speed in rpm,
current in ampere (A).
− T
int
temperature in °C.
T
fix
− Mains voltage in percentage (%) of
nominal and frequency in hertz (Hz).
− Level indicates the liquid level in the
internal bath.
− Voltage of power transformer, 5V supply
and fan in Volt.
− Cur. cons.: Mains current consumption in
Ampere.
Error store
Time
Settings
store
− The last message is at the top.
− Each message line can be marked with
the cursor keys. The message appears in
plain text in the footer.
− Under Source the CAN node is displayed
which signaled the fault.
− Code is the number, which in the Master
is shown in the display until the cause
has been rectified.
T
fix
− Type: Alarm, Warning or Fault (Error).
Maintenance
RgmVW
Settings
Device status
Device data
Display
indicates the current internal bath
Device status
Display
.
Section 7.6.8
Error
115

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rp 855Rp 1290Rp 890Rp 1840Rp 1845Rp 3530 ... Show all

Table of Contents