Integral T process thermostats
7.7.3
Write commands (data commands to the thermostat
Command
OUT_SP_00_XXX.XX
OUT_SP_04_XXX.XX
OUT_SP_05_XXX.XX
OUT_PAR_00_XXX.XX
OUT_PAR_01_XXX
OUT_PAR_04_XXX.XX
OUT_PAR_05_XXX
OUT_PAR_08_XXX.XX
OUT_MODE_00_X
OUT_MODE_01_X
START
STOP
RMP_SELECT_X
RMP_START
RMP_PAUSE
RMP_CONT
RMP_STOP
RMP_RESET
RMP_OUT_00_XXX.XX_XXX
RMP_OUT_02_XXX
21/01/2019 / YAWE 0026
Explanation
Setpoint transfer with max. 3 digits before decimal point and max.
2 decimal digits behind.
TiH outflow temperature high limit.
TiL outflow temperature low limit.
Setting of control parameter Xp for controller (0.1 – 10 °C).
Setting of the control parameter Tn (5 – 200 s).
Setting of control parameter KPE (0.1 – 10.0).
Setting of control parameter TNE (5 – 200 s).
Setting of the WIN value for tolerance range monitoring.
Keys: 0 = free / 1 = inhibited (corresponds to "KEY").
Control: 0=internal / 1 = external.
To switch on the unit (standby modus or "Manual Start" ON if
mains are OFF).
When standby function is on auto (A) there is an error message
(ERR35).
Switches thermostat to standby (pump, heating, refrigeration unit
off). When standby function is on auto (A) there is an error
message (ERR35).
Selection of the programme (1-5) to which the further instructions
apply. When the unit is switched on, programme 5 is selected
automatically.
Start the programmer.
Hold (pause) the programmer.
Restart the programmer after pause.
Terminate the programmer.
Delete the programmer.
Set a programme segment (temperature and time). A segment is
added and appropriate values are applied to it.
Number of times the programme runs: 0 = unlimited / 1 – 250.
Starting up
)
79