Lauda Ecoline E 200 Operating Instructions Manual

Immersion thermostats; bath/ circulation thermostats

Advertisement

Operating Instructions
Ecoline
Immersion thermostats E 200
Bath/ Circulation thermostats
E 206 T, E 212 T, E 215 T, E 220 T
E 203, E 211, E 219, E 225
LAUDA DR. R. WOBSER
GMBH & CO. KG
P.O. Box 1251
97912 Lauda-Königshofen
Germany
From Series Z 01
Phone: (+49) (0) 9343/ 503-0
Fax: (+49) (0) 9343/ 503-222
from software version 3.4
e-mail info@lauda.de
01/02
Internet http://www.lauda.de
YAEE0013

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Ecoline E 200 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lauda Ecoline E 200

  • Page 1 Immersion thermostats E 200 Bath/ Circulation thermostats E 206 T, E 212 T, E 215 T, E 220 T E 203, E 211, E 219, E 225 LAUDA DR. R. WOBSER GMBH & CO. KG P.O. Box 1251 97912 Lauda-Königshofen...
  • Page 3 Safety notes Before operating the equipment please read carefully all the instructions and safety notes. If you have any questions please phone us! Follow the instructions on setting up, operation etc. This is the only way to avoid incorrect operation of the equipment and to ensure full warranty protection. •...
  • Page 4: Table Of Contents

    BRIEF OPERATING INSTRUCTIONS ..................1 CONTROL AND FUNCTIONAL ELEMENTS ................3 UNIT DESCRIPTION........................4 Unit description ......................... 4 Pumps..........................4 Temperature indication, control, and safety circuit ............5 Materials ..........................5 Serial Interfaces RS 232, RS 485 ..................6 3.5.1 RS 232 Interface ..........................6 3.5.2 RS 485 Interface ..........................8...
  • Page 5 6.4.3.3 Selection of the interface ......................29 6.4.3.4 Setting the Baud rate/ transfer rate (serial interface) ..............30 6.4.3.5 User calibration..........................31 Warning and safety functions ..................33 6.5.1 Overtemperature protection and testing ..................33 6.5.2 Low-level protection and testing....................34 6.5.3 Pump motor monitoring ......................35 6.5.4 Connection floating contact ”Combination fault”...
  • Page 6 Explanation of signs Danger: This sign is used where there may be injury to personnel if a recommendation is not followed accurately or is disregarded. Note Here special attention is drawn to some aspect. May include reference to danger. Reference: Refers to other information in different Sections.
  • Page 7: Brief Operating Instructions

    Brief operating instructions Brief operating instructions This brief instruction shall give you the possibility to operate the unit quickly. For safe operation of the unit it is absolutely necessary to read carefully all the instructions and safety notes! 1. Assemble unit and add items as appropriate ( Section 5).
  • Page 8 Brief operating instructions...
  • Page 9: Control And Functional Elements

    Control and functional elements Control and functional elements Clamping bracket Setting knob Pt 100 Pump housing Outlet for bath circulation Jet nozzle Heater Heating LED green; Supply ON Menu functions Mains switch Setting of overtemperature switch-off point Pump connections − inlet Pump connections −...
  • Page 10: Unit Description

    The immersion thermostats E 200 have a device for fixing the immersion thermostat to the bath vessel (clamping bracket). An adapter is supplied for the deep-drawn LAUDA baths 003, 011, 019 and 025. The type designation of the Ecoline bath/circulation thermostats consists of the control unit E 200 and the type of bath.
  • Page 11: Temperature Indication, Control And Safety Circuit

    Unit description Temperature indication, control and safety circuit The unit is provided with a 2-line LCD-Display with additional symbols for indicating bath temperature and settings as well as operating states. The setpoint is input and additional adjustments can be made using either two or three keys.
  • Page 12: Serial Interfaces Rs 232, Rs 485

    Unit description Serial Interfaces RS 232, RS 485 3.5.1 RS 232 Interface Computer Thermostat RS 232 Interface Data 9-pin sub-D 25-pin sub-D 9-pin sub-D socket Data socket socket R x D T x D T x D R x D Signal Signal Ground Ground...
  • Page 13 Unit description Protokoll: − The interface operates with 1 stop bit, no parity bit and 8 data bits. − Transfer rate either 2400, 4800, 9600 (factory setting) or 19200 baud as selected. − The RS232 interface can be operated with or without hardware handshake (RTS/CTS). −...
  • Page 14: Rs 485 Interface

    Unit description 3.5.2 RS 485 Interface Connecting cable: Thermostat 9-pin sub-D-socket Data Data A SG (Signal Ground) optional Data B − Use screened connecting cables. − Connect the screen to the plug case. − The lines are electrically isolated from the remaining electronics. −...
  • Page 15 Unit description − The response of the thermostat is always terminated with CR. − CR = carriage return (hex: 0D). Example: Transfer of setpoint 30.5°C to the thermostat with address 15. Computer Thermostat " „A015_OUT_SP_00_30.5“CR „A015_OK“CR 15.05.01 YAEE0013...
  • Page 16: Write Commands (Data Commands To The Thermostat)

    Unit description 3.5.3 Write commands (data commands to the thermostat) Command Explanation OUT_SP_00_XXX.XX Setpoint transfer with up to 3 places before the decimal point and up to 2 places behind OUT_SP_01_XXX Pump output step 1, 2, 3, 4, or 5; 0 = Stop (Standby) OUT_PAR_00_XXX.XX Setting of control parameter Xp for controller (0.1 –...
  • Page 17: Read Commands (Data Requested From Thermostat)

    Unit description 3.5.4 Read commands (data requested from thermostat) Command Explanation IN_PV_00 Read bath temperature (outflow temperature) IN_PV_01 Read external temperature T IN_SP_00 Read temperature setpoint IN_SP_01 Read pump output step IN_SP_03 Read current overtemperature switch-off point IN_SP_04 Read current outflow temperature limit TiH. IN_SP_05 Read current outflow temperature limit TiL.
  • Page 18: Error Messages

    WK/WKL units can be programmed with the aid of the National Instruments program ® development tool LABVIEW (http://sine.ni.com/apps/we/nioc.vp?cid=1381&lang=US. In order to make program operation possible on the RS 232/ RS 485 interface, LAUDA provides ® drivers specially designed for LABVIEW which can be downloaded free of charge under...
  • Page 19: Unpacking

    Unpacking Unpacking After the unit and accessories have been unpacked they have to be examined for possible transport damage. If there is any damage visible on the unit, the forwarding agent or the post office has to be notified so that the shipment can be examined. Standard accessories: Bath cover E 203 and on all bath/circulation thermostats E 3xx...
  • Page 20: Preparations

    − Adjust the pump chamber. − Move the fixation of the temperature probe downwards again (see ill. on the left) − For all LAUDA baths (plastic and deep-drawn baths), please fix the adapter (standard accessory) on the clamping bracket. − Turn the jet nozzle to face diagonally into the bath. The outflow for the bath circulation can then be closed.
  • Page 21 Preparations Operation with cooling coil Section 10. Accessories) − Pull the clamping bracket to the back for fixing the cooling coil while releasing it with a screwdriver. − Push the cooling coil on the clamping bracket. − Install the clamping bracket again. Operation with fixing rod Section 10.
  • Page 22: Filling And Emptying

    Preparations b) Operation with external consumer (Circulation thermostat) ( Section 5.4.) − At bath temperature above 70°C the label supplied must be affixed on the bath in a clearly visible position. − When operating as bath thermostat without external consumer the pump pressure outflow has to be closed (use closing plugs) or linked to the return.
  • Page 23 Preparations Emptying − Switch off the thermostat, pull out the mains plug! a) Immersion thermostat, Bath/ Circulation thermostat − Unscrew the immersion thermostat or take off the control unit with the bath bridge. − Drain the bath. − The units are designed for operation with non-flammable and flammable liquids to EN 61010-2-010! Flammable liquids can be operated up to no more than 25°C below the firepoint ( Section 5.3.).
  • Page 24: Bath Liquids And Hose Connections

    Preparations Bath liquids and hose connections Bath liquids LAUDA Working Chemical Visco- Viscosity Fire- Ref.No. Designation tem- Designation sity (kin) at point Quantity perature (kin) Temperature range Former from °C at 20°C mm²/s mm²/s 10 l 20 l designation to °C...
  • Page 25 Preparations Hose connections a) Elastomer tubing Tubing type Int. dia. Temperature Application Ref. No. Ø mm range °C EPDM-tubing, 10 to 120 for all bath liquids RKJ 111 uninsulated except for Ultra 350 and mineral oils EPDM-tubing, 10 to 120 for all bath liquids RKJ 112 uninsulated...
  • Page 26: Connection Of External Circuits

    Preparations Connection of external circuits a) Immersion thermostats − Push 11-12 mm int. dia. tubing ( Section 5.3.) directly onto the jet nozzle and connect it to the external circuit. − Hang the return tubing into the bath and fix it! −...
  • Page 27 Preparations b) Bath/Circulation thermostat − Connect 11-12 mm int. dia. tubing ( Section 5.3.) to the pump connector. − Pump outflow connection always in front, return connection always at the back. − If the cross-section of the tubing is too small ! temperature drop between bath and external system due to low flow rate.
  • Page 28: Cooling The Thermostats

    Preparations Cooling the thermostats At bath temperatures down to just above ambient temperature (approx. 2 – 10°C) it is possible to work without cooling. Additional cooling is required for lower temperatures ! attach the cooling coil ( Immersion thermostat. Section 5.1.). Bath/Circulation thermostats ! fitted with cooling coil, as standard..
  • Page 29: Starting Up

    Starting up Starting up Connection to the supply Compare the supply voltage against the data on the rating label. Model according to EMC directive EN 61326-1 Class B.* − Connect the unit only to a grounded mains power socket (PE). −...
  • Page 30: Setpoint Selection (Level 0)

    Starting up $ $ $ $ − A double signal tone sounds. − The display for low-level (LEVEL) appears. − The fault triangle is flashing − Press the key. If necessary repeat several times. − Also press the key if the unit had switched off under a fault condition.
  • Page 31: Menu Functions

    Starting up Menu functions − Switching from setpoint selection (level 0) to level 1 using the key − Within one level it is possible to scroll using the keys. − In principle, after each setting has been made it is entered automatically after approx.
  • Page 32: Setting The Setpoint Resolution

    Starting up 6.4.2 Setting the setpoint resolution − To set the setpoint resolution from level 0 press the key and 1x combination on the left or − move forward with the key from the PUMP-function. − The current bath temperature, RES and the current setpoint resolution are indicated.
  • Page 33: Parameters

    Starting up 6.4.3 Parameters − Directly from level 0 (setpoint selection) press the key and 2x combination on the left, or − forward with the key from setpoint resolution function. − Here it is possible to switch over to level 2. Press the key on the left, continue with Section 6.4.3.1.
  • Page 34: Setting The Reset Time Of The Pid-Controller

    Starting up − The display shows the current bath temperature Xp and the current setting. To alter the setting press the key on the left. Available settings from 0,5 to 9,9 °C . Section 6.4.3.2. Note). Section 9. Technical data) −...
  • Page 35: Selection Of The Interface

    Starting up − During this time, start to set the required value with one of the two keys. − Forward with the key to "Selection of the interface" or − with key back to "Setting the proportional band“. − The control parameters are pre-set to suit the unit type. Normally no change is required.
  • Page 36: Setting The Baud Rate/ Transfer Rate (Serial Interface)

    Starting up − The display shows the current bath temperature, ADR and the current unit address (e.g.. 15). Press the key, display flashes (approx. 4 s.). − During this time start to set the required unit address. (possible setting: 000...127) −...
  • Page 37: User Calibration

    Starting up − with key back to "Selection of the interface". 6.4.3.5 User calibration − Remove the external consumers and switch the setting knob of the pump to right side. ( Section 5.4). − A reference thermometer with necessary accuracy is required. Otherwise the factory calibration should not be altered.
  • Page 38 Starting up 5. Forward with key to "END", then 6. Switch back to level 0 or 7. with key back to "Setting the transfer rate". Example a) Insert a suitable thermometer into the bath (long enough and far enough). b) Remove the external consumers and turn the setting knob of the pump outflows to the right side.
  • Page 39: Warning And Safety Functions

    Starting up Warning and safety functions 6.5.1 Overtemperature protection and testing − The units are designed for operation with non-flammable and flammable liquids to EN 61010-2-010! − Set the overtemperature switch-off point. Recommended setting 5°C above required bath temperature. − Not higher than 25 °C below the firepoint of the bath liquid ( Section 5.3.).
  • Page 40: Low-Level Protection And Testing

    ! − If there is any irregularity when testing the safety devices, switch off the unit immediately and pull out the mains plug ! − Have the unit checked by the LAUDA service or the local service organisation!
  • Page 41: Pump Motor Monitoring

    Starting up 6.5.3 Pump motor monitoring − In case of pump motor overload or a blockage the heating and the pump are switched off. ! ! ! ! − Double signal tone sounds. − The display shows PUMP and the fault triangle is flashing −...
  • Page 42: Other Error Messages

    Starting up 6.5.5 Other error messages − E 0 etc. is flashing in the bottom line. ! various temperature probe faults. ! pump fault, proceed as in Section 6.5.3. − If the fault report is repeated ! pull out the mains plug and try whether the motor can be rotated by the fan blade inserting a screwdriver into the ventilation opening at the back of the unit.
  • Page 43: Safety Notes

    Safety notes Safety notes General safety notes A laboratory thermostat is intended for heating and pumping liquids according to the needs of the user. This leads to hazards by high temperatures, fire, and the general hazards by the use of electrical energy. The user is largely protected through the application of the appropriate standard specifications.
  • Page 44 Safety notes produced. Ensure appropriate ventilation! • Immersion thermostats have to be fixed carefully at the bath vessels! • Only use bath vessels which are appropriate for the intended operating temperatures! • When changing the bath liquid from water to oil, for temperatures above 100 °C, carefully remove all traces of water, also from tubing and from the external consumer, otherwise ! danger of burns through delayed boiling! •...
  • Page 45: Maintenance

    − Before any maintenance and repair work pull out the mains plug! − Repair on the control unit must only be carried out by properly qualified personnel! LAUDA thermostats are largely maintenance-free. If the thermostating liquid becomes dirty it has to be replaced ( Section 5.2.).
  • Page 46: Ordering Spares

    When ordering spares please quote instrument type and serial number from the rating label. This avoids queries and supply of incorrect items. We shall always be happy to deal with queries and to receive suggestions and criticism LAUDA DR. R. WOBSER GMBH & CO.KG P.O. Box 1251 97912 Lauda-Königshofen...
  • Page 47: Technical Data(To Din 58966)

    Technical data Technical data (to DIN 58966) Common technical data E 200 Ambient temperature range °C 5 to 40 Setting resolution °C 0.1/ 0.01 Indication resolution °C 0.05 Indication accuracy °C ± 0,2 °C additive re-calibration Temperature control ± °C 0,01 Safety features Additional functions...
  • Page 48 Technical data E 203 E 211 E 219 E 225 Operating temperature range 23 to 150 “with water cooling °C 20 to 150 Operating temperature range °C -20 to 150 Heater 230 V; 50/60 Hz 2.25 power 115 V; 60 Hz 100 V;...
  • Page 49 Technical data Equipment types Xp ( °C ) Tn ( s ) Pump output step E 200 E 203 E 211, E 212 T E 206 T E 215 T E 219 E 225, E 220 T We reserve the right to make technical alterations! 15.05.01 YAEE0013...
  • Page 50 Technical data Pump characteristics Immersion thermostats measured with water 0,45 0,35 step 5 step 4 0,25 step 3 step 2 0,15 step 1 0,05 l/min Bath/ Circulation thermostats measured with water 0,45 0,35 step 5 step 4 0,25 step 3 step 2 0,15 step 1...
  • Page 51: Accessories

    Accessories Accessories Immersion thermostats Accessories Ref. No. Cooling coil HOK 064 Pump set (outflow and return connection) LCZ 0638 Fixing rod LCZ 0637 Bath Materials Max. Temp Volume (l) Inner size (WxDxH) Ref. No. (°C) 006 T polycarbonate 5 to 7 130x420x160 * LCZ 0628 015 T...
  • Page 52 Accessories Bath / Circulation thermostats Accessories suitable for Ref. No. Bath cover E 211 HDQ 079 Bath cover two parts E 219, E 225 LCZ 0632 Gable cover E 220 T LCZ 011 Gable cover E 219, E 225 LCZ 0634 Cover plate MD 15 K E 215 T LSZ 0115...
  • Page 53: Circuit Diagram

    Circuit diagram Circuit diagram 14.04.99 YAEE0013...
  • Page 54 Circuit diagram 230V; 50Hz 230V; 50/60Hz [230V; 60Hz] at serial no: X01 E 1xx E 2xx Printed circuit board „Mains“ UL 487-1 UL 487-1 Printed circuit board „Display“ UL 488-1A UL 488-1B Printed circuit board serial interface RS 232/RS 485 ---------- UL 490 Printed circuit board Mains LED-Backlight...
  • Page 55 / Personne responsable Hiermit bestätigen wir, daß nachfolgend aufgeführtes LAUDA-Gerät (Daten vom Typenschild): We herewith confirm that the following LAUDA-equipment (see label): Par la présente nous confirmons que l’appareil LAUDA (voir plaque signalétique): Serien-Nr. Type / Type Serial no. / No. de série:...

Table of Contents

Save PDF