9.2.4 Device data
Master
Command
T ext Pt
T ext analog
T ext serial
T cont. head 39.80 Level
T heatsink
Pump Pow.
Pump rpm
Pumpe Cur.
Pump
Menu
9.2.5 Fault memory
For the analysis and localization of faults the Command version includes a fault memory in which up to
45 fault and alarm messages are saved.
Command
No Source Code Type
10 Safety 2
9 Safety 4
8 Contro. 32
7 Contro. 3
6 Contro. 9
5 Contro. 3
4 Contro. 4
3 Contro. 5
Low level
Pump
Menu
100
25.70 Tint
---.-- Mains U(%) 103.74
---.-- Mains Frequ.
51.68 Low voltage 27.90
44.90 5V Supply
5460 Fan Voltage
1.68 Cur. cons.
End
T
set
Date
Alarm ------------ -----------
Warn. 28.08.03 15:32:02
Error 17.07.03 10:52:02
Warn. 06.06.03 11:15:11
Alarm 05.06.03 08:45:01
Alarm 01.06.03 17:58:22
Warn. 28.05.03 20:01:22
Warn. 27.05.03 07:58:00
End
T
set
Maintenance
Proline Edition X heating thermostats
MNEmv ShoWX (
Various device datas are displayed.
Device data
25,58
vice data
50
Text shows various actual temperatures
6
in °C from ext. Pt100 and the modules.
T cont. head and T heatsink are tempera-
5.00
tures of electronics in the Master in °C.
7.0
Pump power in Watts, pump speed in
2,84
rpm, pump current in ampere.
T
int
temperature in °C.
Mains voltage in % of nominal and mains
T
fix
frequency in hertz.
Level indicates the liquid level in the in-
ternal bath.
Low Voltage of power transformer, of the
5 V supply and fan voltage in volt.
Cur. cons.: Mains current consumption in
ampere.
Errorstore
Time
Settings
store
The last message is at the top.
Each message line can be marked with
the cursor keys. The message appears in
plain text in the footer.
Under "Source", the CAN node is dis-
played which signaled the fault.
Code is the number, which in the Master
is shown in the display until the cause
has been rectified.
T
fix
Type: Alarm, Warning or Fault (Error).
Section 7.5)
Settings
Device status
Display
.
indicates the current internal bath
Device status
Display
.
YACE0099
De-
Error-
/