4.3 Calibration Wizard

For the user's convenience a TTC/12TC-wizard is included in the software.
To start the wizard, please use the magic wand from the Main Window's
toolbar or select Extras
selection dialog appears and lets the user choose between:
1.
Differential TTC (DTTC, available with differential measurement mod-
ules)
2.
Single-ended TTC (For channel 1 only. Use DTTC for channel 2)
3.
Save current for TTC/12TC for immediate reload
4.
Reload last known TTC/12TC
5.
Reset currently loaded TTC/12TC
6.
Full 2-port calibration (only if F2P module is available)
If you choose 1., 2., or
sary steps towards a reliable TTC, DTTC or 12TC; nevertheless subsequently
some helpful remarks are enlisted:
In the start window, the wizard allows for reloading an existing TTC/12TC.
Please ensure that the currently used measurement setup (in the first
place the attached cables and time base settings of the device) is iden-
tical to that of the original TTC/12TC.
If a TTC/12TC is loaded from a previously wizard-saved SQC-file, the
wizard jumps directly to the last window and you are done. This is
only necessary if different TTCs/12TCs (e.g. for different time bases)
shall be handled in parallel.
Three-Term Correction from the main menu. A
6.
the TTC-wizard will guide you through the neces-
T3SP-Series Operator's Manual
31
931776 RevB