Teledyne T3SP Series Operator's Manual page 21

Time domain reflectometer
Hide thumbs Also See for T3SP Series:
Table of Contents

Advertisement

The precision of the cursor values can be changed via the menu item View
Set decimal places of cursor.
Please note: at most two cursors can be set for each sample. That means that if two
cursors are already set in impedance measurement mode and the measurement
mode is changed (e.g. to S11 mode), no more cursors can be set. In this case delete
the existing cursors before setting them in S11 mode.
Legend
By default the legend shows the sample name (only the first 8 characters).
Alternatively the samples row number corresponding to the spread sheet
row can be displayed. Use View
between the modes.
Data Export
The underlying data of the currently displayed samples can be exported to
various formats. Currently supported formats are csv, xlsx, mat (Mat-
lab) and sp1/sp2 (Touchstone for S-parameters). To export data via the
Result Visualization Window use the export button or the menu entry File
Export.
Please note: the Touchstone export is available for S-parameter modes only. Data
can be saved to the Touchstone format. Each selected sample is exported into an
individual file. The frequency data points are exported in MHz, the scattering pa-
rameter values as magnitudes and degrees. Optionally the magnitude of the data
can be limited to 0dB.
Background Color
The background color of the diagram can be changed via the menu entry
View
Change Background Color.
T3SP-Series Operator's Manual
Show sample names in legend to toggle
21
931776 RevE