Teledyne T3SP Series Operator's Manual page 93

Time domain reflectometer
Hide thumbs Also See for T3SP Series:
Table of Contents

Advertisement

No calibration kit is
available for performing the
three-term calibration
(open, short, load, thru)
When pressing the
measurement button, the
message "Temperature not
stable" appears.
T3SP-Series Operator's Manual
No ckd file containing calibration kit
coefficients can be found by the software.
Make sure it is copied to the programs
Config subfolder. Alternatively it can be
installed via the keyboard shortcut CTRL-I
or the menu entry File
Main Window.
The internal device temperature must be
stable for ensuring the internal
temperature compensation to work
properly. This message usally appears at
startup or if the device is exposed to
extreme temperature changes.
Install in the
93
931776 RevE