2.5.3 Result Visualization Window

Figure 3: Impedance measurements results
General Remark
Most Measurement Modes (see chapter 5) yield more information than can
be displayed in the yellow result field of the Main Window. To display the
specific results in more detail the Result Visualization Window is used. The
different kinds of visualizations depend on the currently selected measure-
ment mode and are therefore described in the specific sections.
TTC/12TC Calibration
The software allows the correction of systematic errors by means of OPEN,
SHORT, LOAD calibration, referred to here as Three Term Correction (TTC).