T3SP-Series Operator's Manual
Figure 5: Result visualization of impedance measurements
tolerance check failed. Additional LEDs (green/red) are located in the
status bar of the Result Visualization Window (only if exactly one sam-
ple selected in list).
Please note: two different options can be chosen to classify the measure-
ment results. In the strict mode, the measurement is classified as NOT PASSED
if the tolerance mask is violated at any point. In the non-strict mode, the
measurement is classified as NOT PASSED if the average impedance result is
out of the limits. The mode can be set in the Seunis.config file (see section
2.2) by setting option
Pass-fail strict either to "1" or "0" .
The tolerance mask can be turned on/off by using the
button in
the tool bar. This also effects displayed impedance results (average &
min./max. impedance) and the small LED indicators in the status bar.
The indicators are grayed out if the tolerance mask is switched off.