Teledyne T3SP Series Operator's Manual page 62

Time domain reflectometer
Hide thumbs Also See for T3SP Series:
Table of Contents

Advertisement

can be found in toolbar of the Main Window. If no de-embedding param-
eters are loaded, the button is grey and becomes green only after the S-
parameters are loaded.
The S-parameters to be loaded must be available in Touchstone (version 1
or 2) files, which can either be s2p (2-port) or s4p (4-port) files. In case a
s4p file is provided, the user has to take care of selecting the correct ports,
enabling the software to extract the desired set of 2-port parameters from
the Touchstone file.
For loading the fixtures S-parameters please proceed as follows:
1.
Open the De-Embedding Dialog.
2.
Select the Touchstone file for channel 1 and 2. In case a s4p is se-
lected, a user input dialog for the ports association opens.
3.
Press the Apply or the OK button. Pressing OK applies the parameters
and closes the dialog immediately.
4.
The de-embedding of the individual channels can be (de)activated by
setting the corresponding checkboxes.
Figure 12: De-Embedding Dialog
T3SP-Series Operator's Manual
62
931776 RevE