at this stage.
•
On completion of the wizard the TTC/12TC data is automatically stored
for reuse (see 4.: Reload last known). The previous TTC/12TC is re-
placed and cannot be recovered anymore.
•
A differential TTC (DTTC) can be extended to a 12TC. If a DTTC is active
and a full 2-port calibration is initiated, the wizard asks the user, if the
current DTTC shall be extended.
•
In case a multiplexer is connected, please refer to chapter
ther information about calibrating multiple channels.
4.4 General Remarks on reloading TTC/12TC Data
•
The last known TTC/12TC is loaded automatically on program startup,
if activated in the Options Window
•
The last available TTC/12TC can easily be reloaded by opening the TTC-
Wizard and pressing the Reload button.
•
The software shows a warning, if the loaded TTC/12TC is older than
14 days. This value can be changed in the Seunis.config (see section
2.2) file by editing the Old TTC warning after <N> days entry.
•
TTC data is always stored and reloaded with project files. Please be
aware that after loading an older file containing valid TTC/12TC-data,
the currently active TTC/12TC is replaced.
Remark: this behavior is essential for the correct recalculation of the results
of the stored data and is explicitly an intended behavior. An option to circum-
vent this, is to work with workshop files (see chapter 9).
•
The user is not allowed to change time base settings when using a
T3SP-Series Operator's Manual
Device tab.