Before a maintenance work, be sure to carry out the setting explained in this subsection to hold the analog
output.
Select "Output Hold" from the "Menu" screen, and press the

6.4.1 Output hold

Set output hold to "Last Meas." or "Pre-set" to hold analog output. (Indication value on the "Measurement"
screen is not held. However, it is highlighted during the output hold time.)
Output Hold
HCℓ
Output Hold
HCℓ
6.4 Output hold
2018-04-01
12:00:00
No
020
%FS
(
)
2018-04-01
12:00:00
Pre-set
020
%FS
(
)
NT
NT
key.
Point the
to the measurable component to
hold output by the
and press the
Select either "Last Meas." or "Pre-set" by the
key or the
to validate the setting.