Setting Change Items; Zero Limit - Fuji Electric ZPA Service Manual

Infrared gas analyzer
Hide thumbs Also See for ZPA:
Table of Contents

Advertisement

(2) Setting change items

1) Zero limit

Function: Switches measured concentration values below zero to either display or no display mode.
Operation method:
Set values are inverted by pressing the
when the cursor is aligned with the "Zero limit".
Switch between "ON" and "OFF" by the
Establish the setting contents by pressing the
If you don't want to establish the contents, press the
key.
Setting contents:
OFF: does not display and output values below zero
ON: displays and outputs values below zero.
Initial value: "OFF" (default: Disappear)
* This mode is used at the time of adjustment in order to check a display. If the minus display (Zero
limit) is set to Appear, be sure to return the setting to "OFF" after adjustment.
TN5A1191-E
Changes the setting in "Factory mode", "12. Others" and "Zero limit".
ENT
Factory mode initial screen
The cursor is in 12.
key,
or
key.
ENT
key.
ESC
ENT
ESC
ENT
ESC
21