6.3

During use (When alarm is generated)

LCD display when some alarms are generated
*See Chapter 4 for operation of buzzer, vibration and LED/Backlight during alarm generation.
No
Item
1
Accumulated dose
alarm/warning
2
Dose rate alarm/warning
3
Operation time alarm
4
Low battery voltage
Display
Alarm
Overload
Warning
Alarm
Overload
Warning
Remarks
Warning alarm is
generated when
exceeding dose
warning set value.
Alarm is generated
when exceeding dose
alarm set value.
Also, display is
changed when
overload happens.
Warning alarm is
generated when
exceeding dose rate
warning set value.
Alarm is generated
when exceeding dose
rate alarm set value.
Also, display is
changed when
overload happens.
Alarm is canceled when
dose rate reaches
down at 80% of
warning/alarm value or
less.
When exceeding
operating time alarm
set value, time alarm is
generated.
Low battery voltage
alarm is generated
when battery voltage
reaches less than 1.1V.
Remaining operational
hours is displayed with
numbers 1 to 9 h, then
power OFF after 1 hour.