Installation Procedure - Thermo Scientific Smart-Vue Installation Manual

:luhohvv6luhq
Hide thumbs Also See for Smart-Vue:
Table of Contents

Advertisement

Installation procedure

Placing your Thermo
Scientific siren for best
wireless performance
Thermo Fisher Scientific
The procedure described here assumes that Smart-Vue Server/ Smart-Vue
Client software is installed on your computer, and that your Thermo
Scientific wireless receiver is configured and ready to use.
For optimal operation, follow these recommendations when physically
placing your device:
Do not place the siren within 40 cm (16 in.) of another end-point
module or wireless device.
Make sure the wireless end-point module is not placed on an electrical
conduit or cable tray (such as for computer network cables).
For best results, place the end-point module so that it faces the general
direction of the receiver antenna if possible.
Keep about 20 cm (8 in.) of clear space around the end-point module.
For example, a end-point module that is "stuck" between two
refrigerators may not communicate effectively.
Place your siren in a relatively high location on a wall where people will
not bump or unplug it (e.g., about 2 meters (6.5 ft.) from the floor).
To ensure that the AC adapter does not get unplugged accidently, use a
power socket that is also elevated from the floor and/or shield the cord
in a protective conduit that is fixed to the wall (or within the wall).
The siren is a critical device in your wireless monitoring system.
Therefore, we recommend that you keep it plugged it into an
Uninterruptible Power Supply or safety plug.
See Appendix ref on pg.14 for additional considerations.
Installation procedure
Smart-Vue Wireless Siren Installation Guide
8