Installation Environmental
Considerations
The instrument is designed for general-duty, indoor
installations. The electronics enclosure environmental rating is
NEMA 12 and the Fluidics enclosure is a NEMA 4X fiberglass
with a clear door window.
The Orion 2210AM Ammonia Analyzer should be installed in a
well-ventilated location, with ambient temperature between
5°C and 45°C (41°F to 113°F); and relative humidity not to
exceed 90% at 40°C. The area must be free from dust,
corrosive gases, vibration and shocks; sheltered from direct
sunlight, and shielded from dripping water. Power should be
instrument grade with over voltage protection, and a provision
for disconnecting power.

Mounting the Instrument

The instrument case is designed for wall or panel mounting.
Use four #10 screws or equivalent for proper support.

Refer to Figure 3 and Figure 4 for dimensions (in mm and

inches).
Make sure that the wall that you choose can support the

weight of the instrument (approximately 20 kg or 44 lbs.)
(See Figure 3).
Locate the instrument as close to the sampling point as

practical to ensure complete purging of the sample line
during each cycle.
Allow approximately 50 cm (20 in.) of clearance at the left

side for door swing. Also allow enough clearance on the
right side for wiring conduit.
Locate the analyzer height such that the display is at

eye-level height for an operator, allowing also for
appropriate sample and drain connections.
Orion 2210AM Ammonia Analyzer
Figure 3. Mounting dimensions

Installation Environmental Considerations

|
9