Starting the detector

Warning:
damage the instrument. See
more information.
Starting the detector entails powering-on the detector and each system
instrument individually, as well as the ACQUITY workstation. It also entails
starting the operating software (Empower, MassLynx, or third party).
Caution:
• To ensure a long life for the light-guiding flow cell and proper
detector initialization, use well-degassed eluents, making sure they
are flowing before you power-on the detector.
• To minimize contaminants that can leave deposits on the flow cell's
walls, flush new columns for 15 minutes before connecting the flow
cell.
If you must power-on the detector before the eluent is flowing, extinguish the
lamp. You can do this in the Instrument Method Editor (Empower, MassLynx,
or third-party) by specifying a Lamp Off event in the Events table. You can
also extinguish the lamp in one of these ways:
If Empower software controls the system, click
control panel at the bottom of the Run Samples window.
If MassLynx software controls the system, click
control panel at the bottom of the Inlet Editor window.
In the console, select PDA/eλPDA Detector from the system tree and
click
ACQUITY UPLC System Documentation CD or ACQUITY UPLC
See also:
H-Class System Documentation CD.
To start the detector
1.
Power-on the workstation.
2.
Press the power switch on the top, left side of the solvent manager
(QSM/BSM) door and sample manager door.
Using incompatible solvents can injure you and severely
(Lamp Off).
"Solvent considerations" on page 98
Starting the detector
for
(Lamp Off) in the
(Lamp Off) in the
29